51
NT UF 811 CP GB (e) / June 2015
The relays may be tested individually: select the open or closed mode on the relay state line.
5.4.5 "Logger settings" menu (if enabled)
Important:
The logger must be reset before the number of variables is
changed.
1) Resetting the logger
To reset the logger, enter "YES" in the "RESET LOGGER" field.
2) Confirmation of resetting the logger
Resetting the logger requires the user to enter a confirmation field (protection against handling errors).
3) Number of parameters for the logger
Up to 30 variables can be recorded in the logger. The number of variables can be adjusted. Its maximum
autonomy is for 530,000 time-stamped readings.
4) Logger mode
The logger has 2 recording modes: "rotating" (initial files overwritten when the logger is full) or "Stop when
full” (the logger stops operating when it is full).
5) Logger interval
The recording period of the logger can be set from 1s to 24 hrs: 1 s, 5 s, 10 s, 30 s, 1 min, 2 mins, 5 mins,
10 mins, 15 mins, 30 mins, 1 hr, 2 hrs, 6 hrs, 12 hrs and 24 hrs.
6) Logger autonomy
This field, which cannot be changed, shows the remaining autonomy of the logger.
Содержание Uf 811
Страница 4: ...4 NT UF 811 CP GB e June 2015 CHAPTER 1 OVERVIEW...
Страница 11: ...11 NT UF 811 CP GB e June 2015 CHAPTER 2 SAFETY INSTRUCTIONS...
Страница 14: ...14 NT UF 811 CP GB e June 2015 CHAPTER 3 INSTALLATION AND WIRING...
Страница 28: ...28 NT UF 811 CP GB e June 2015 CHAPTER 4 INSTALLING A MEASURING POINT...
Страница 30: ...30 NT UF 811 CP GB e June 2015...
Страница 38: ...38 NT UF 811 CP GB e June 2015 CHAPTER 5 USING AND CONFIGURING THE Uf 811...
Страница 65: ...65 NT UF 811 CP GB e June 2015 CHAPTER 6 FUNCTION ENGINE...
Страница 67: ...67 NT UF 811 CP GB e June 2015 CHAPTER 7 PC SOFTWARE...
Страница 81: ...81 NT UF 811 CP GB e June 2015 Appendix I Characteristic echo signals...
Страница 84: ...84 NT UF 811 CP GB e June 2015 Appendix II Speed of sound in water...