The detection wavelength range is manually selected by using a 0.5” bandpass filter. The labeled
broad rectangular slit on top of the Kronos housing accommodates one such filter mount. Figure 6
shows the transmission spectrum of the 600 nm bandpass filter included with Kronos. These filters
have a bandwidth, or full width at half maximum of 10 nm (in this scenario, 595 nm to 605 nm). Any
incident light within this bandwidth has high transmittance.
Figure 6:
Transmission spectrum of 600 nm detection dielectric interference filter.
7.4 Acquiring Data
Place a filled cuvette into the sample area and select the appropriate parameters for averaging and
time window. For absorption measurements, click on the Absorption radio button. Ensure the
Range is set to auto and do not disable this for absorption measurements. The software will
automatically adjust the detector gain and LED intensity. In absorption mode, both LED and Xe light
sources will be activated.
18