5. Maintenance
Repair
User-Performed Repair
1. Routinely check the lens for cured dental resins. If necessary, use a non-diamond dental instrument to carefully remove any adhered resin.
2. Light meters differ greatly and are designed for specific light guide tips and lenses. Ultradent recommends routinely checking the output in Standard Power mode. NOTE: the true
numeric output will be skewed due the inaccuracy of common light meters and the custom LED pack in the curing light.
Manufacturer Repair
1. Repairs are only to be performed by authorized service personnel. Ultradent to provide service personnel with documentation to perform repairs.
Warranty
Ultradent hereby warrants that this instrument shall, for a period of 5 years*, conform in all material respects to the specifications therefore as set forth in Ultradent’s documentation
accompanying the product and be free from any defects in materials/or workmanship. This warranty applies solely to the original purchaser and is not transferable. All defective products
are to be returned to Ultradent. There are no user service components of the VALO system. Tampering with VALO will void its warranty.
The VALO warranty does not cover customer damage. For example; if a VALO is misused or dropped and the lens breaks, the customer would be responsible to pay for any necessary
repairs.
*With sales receipt indicating the date of sale to the dentist.
6. Processing
After each use, moisten a gauze or soft cloth with an approved surface disinfectant and wipe the surface and lens.
ACCEPTABLE CLEANERS:
• Lysol Brand III Disinfectant Spray (Recommended)
• Isopropyl alcohol
• Ethyl alcohol-based cleaners
• Lysol®* Concentrate (alcohol-based only)
• Cavicide™* products (Non-Bleach)**
UNACCEPTABLE CLEARNERS - DO NOT USE:
• Strong alkali detergent of any type, including hand soaps and dish soaps
• Bleach-based cleaners (e.g. Clorox™*, Sterilox™*)
• Hydrogen Peroxide based cleaners
• Abrasive Cleansers (e.g. Comet Cleanser™*)
• Acetone or hydrocarbon-based cleaners
• MEK (Methyl Ethyl Ketone)
• Birex®*
• Gluteraldehyde
• Quaternary Ammonium Chloride salt-based cleaners (except Cavicide™*)
• Cavicide1™* solution or wipes
*Trademark of a company other than Ultradent
** If used, it may fade the color
7. Storage and Disposal
Curing light Storage and Transport:
• Temperature: +10°C to +40°C (+50°F to +104°F)
• Relative Humidity: 10% to 95%
• Ambient Pressure: 500 hPa to 1060 hPa
When disposing of electronic waste, (i.e. devices, chargers, batteries and power supplies), follow local waste and recycling guidelines.