8 Installation and Set-Up
8.1 Setting up mattress
Secure the mattress to the moving parts of the bedframe using the securing straps, and when the
system is fully inflated it is ready for the patient to be positioned onto the mattress. The mattress is
designed to completely replace any existing mattress which may be in use on a bed.
Remove any existing mattress and ensure that there are no protruding parts or sharp objects on the
bed which could cause damage to the mattress. Lay the mattress on the bed patient surface ensuring
that the air pipe connector is situated on the bottom left hand side of the bed foot end (as viewed from
the foot of the bed). There are security straps fitted to the base cover of the mattress which should be
fastened loosely to convenient points on the bedframe patient surface.
Caution: Ensure when fixing the mattress to the bed the security straps are only
fitted to the moving parts of the mattress platform. Straps secured to the fixed parts
of the mattress platform will damage the bed/mattress when operated.
8.2 Setting up the Pump
Whilst holding the pump, unfold the hanging hooks on the rear of the Pump and hang it from the bed
’s
footboard. If required extra wide hooks may be purchased to enable the pump to hang on very wide
beds.
Caution: Ensure that the power cable and the air output pipe are routed in such a
manner so that they cannot be twisted, trapped, crushed or stressed.
Pump Unit
Mattress
CPR Valve
Air Pipe Connector
HEAD END
FOOT END