Tamora III Advance
Quick Reference Guide
Please refer to the product User Manual for full instructions for use.
Control Panel Lock Out
To unlock panel:
•
Press & hold for 3 seconds to unlock control panel.
To lock panel:
•
To lock out the control panel press the
Lock Out
button and all functions will be locked.
•
If the control panel is not used for a period of 30
seconds it will automatically lock out and inhibit
the use of the functions.
Alternating Mode
Static Mode
•
Pressing the
Mode
button until the
Static
indicator
illuminates puts the system into
Static
mode.
•
Th
e system will automatically
r
eturn to
Alternating
mode
after 20
minutes.
Alternating Cycle Time selection
•
Choice of
10/
15/20/25/30 minutes.
•
To select press
Cycle
button until required cycle
time is displayed in LCD screen.
Auto Firm Mode
•
•
Pressing the
Auto Firm
button quickly inflates
all
cells to a single pressure setting.
The system will automatically return to
Last Mode of Operation
after 20 minutes.
Comfort Control
The
Soft
and
Firm
buttons allow adjustment of
pressures within a safe pre-set range.
•
When pressing the
Firm
button, the output
pressure
will increase to provide a higher
pressure output
and thus increased support.
•
When pressing the
Soft
button, the output
pressure will be decreased to provide a lower
pressure output and thus increased comfort.
•
Alternating
mode is the default mode for the
and deflate in an alternating 1-in-3 cell cycle.
Function Mode Button
•
Pressing the
Mode
button allows you
to cycle
between the different modes
available such as
Alternating, Static
and
Constant Low Pressure
.
Alarm
s
Alarm Mute
The
Alarm Mute
button temporarily resets the
audible
Low Pressure
/
Power Failure
/
Service
alarms. Should the situation not be resolved and
the fault condition continues the alarm will resume.
Low Pressure Alarm
•
Upon detection of low pressure, an audible alarm will be
heard and the
Low Pressure
indicator will illuminate.
Power Failure Alarm
•
If mains power is removed from power unit or the power
cable is
unplugged without turning the power unit OFF, an
audible alarm
will sound and the
Power Failure
indicator
will
illuminate.
Service Alarm
•
Should your system develop a fault condition whilst in use,
an
audible alarm will be heard and the
Service
indicator will
illuminate.
Constant Low Pressure (CLP) Mode
www.ultimatehealthcare.co.uk | Tel: 0333 321 8996
Auto-Detection
-
detecting patient weight (distribution)
•
Pressing the
Mode
button until the
CLP
indicator
illuminates puts the system into
Constant Low
Pressure
mode.
•
Within this mode the cells are not alternating and the
internal cell pressure within the mattress is set at 15mmhg
to provide a comfortable pressure reducing surface.
•
The
Alternating
cycle will continue at the selected
cycle time until another mode is selected.
•
•
Simultaneously pressing the
Soft
and
Firm
buttons on the
control panel will manually activate
Auto-Detection
.
2QFHDFWLYDWHGWKHLQGLFDWRUOLJKWZLOOIODVKDQGWKHV\VWHP
will start detecting the patients weight
GLVWULEXWLRQ
to set
internal pressures
appropriately.
The system will
continuously check this every 2 hours.