Heritage II Digital Turn
Quick Reference Guide:
Please refer to the product User Manual for full instructions for use.
Alternating Mode
Control Panel Lock Out
To unlock panel:
•
Press & hold for 3 seconds to unlock control panel.
To lock panel:
•
To lock out the control panel press the
Lock Out
button and all functions will be locked.
•
If the control panel is not used for a period of 30
seconds it will automatically lock out and inhibit
the use of the functions.
•
Alternating
mode is the default mode for the
V\VWHP:LWKLQWKLVPRGHWKHPDWWUHVVZLOOLQÀDWH
DQGGHÀDWHLQDQDOWHUQDWLQJLQFHOOF\FOH
•
The alternating cycle will continue at the selected
cycle time until another mode is selected.
Static Mode
•
Pressing the
Mode
button until the
Static
indicator
illuminates puts the system into
Static
mode.
•
The system will automatically return to
Alternating mode
after 20 minutes.
Alternating Cycle Time Selection
•
&KRLFHRIPLQXWHV
•
To select press
Cycle
button until required cycle
time is displayed in LCD screen.
Auto Firm Mode
•
Pressing the
Auto Firm
EXWWRQ
will inflate cells
to the
Max Weight Banding
in order to create a
firm surface
•
The system will automatically return to
Last Mode
Of Operation
after
20 minutes.
Comfort Control
The
Soft
and
Firm
buttons allow adjustment of
SUHVVXUHVZLWKLQDVDIHSUHVHWUDQJH
•
When pressing the
Firm
button, the output pressure
will increase to provide a higher pressure output
and thus increased support.
•
When pressing the Soft button, the output
pressure will be decreased to provide a lower
pressure output and thus increased comfort.
&RQVWDQW/RZ3UHVVXUH&/3
Mode
•
Pressing the
Mode
button until the CLP indicator
illuminates puts the system into
Constant Low
Pressure
mode.
•
:LWKLQWKLVPRGHWKHFHOOVDUHQRWDOWHUQDWLQJDQG
WKHLQWHUQDOFHOOSUHVVXUHZLWKLQWKHPDWWUHVVLVVHW
DWPPKJWRSURYLGHDFRPIRUWDEOHSUHVVXUH
UHGXFLQJVXUIDFH
•
•
Simultaneously pressing the
Soft
and
Firm
buttons on the
control panel will manually activate
Auto-Detection
.
2QFHDFWLYDWHGWKHLQGLFDWRUOLJKWZLOOIODVKDQGWKHV\VWHP
will start detecting the patients weight
GLVWULEXWLRQ
to set
internal pressures
appropriately.
The system will
continuously check this every 2 hours.
Auto-Detection -
detecting patient weight (distribution)
30
130
255
50 75 100
165 205
kg
CLP
Alarms
Mute
The
Alarm Mute
button temporarily resets
the
audible
Low Pressure
/
Power Failure
/
Service
alarms. Should the situation not be resolved and
the fault condition continues the alarm will resume.
Please refer to the manual
for other alarms
Mode
Function Mode button
•
Pressing the
Mode
button allows you to cycle
between
the different modes available such as
Alternating
Static
DQG
Constant Low Pressure
A Full
R
isk
A
ssessment should be completed before using this System
-
Please refer to User Manual for guidance
www.ultimatehealthcare.co.uk | Tel: 0333 321 8996
In the event of a patient suffering a cardiac
arrest and CPR needing to be administered:
for
up to 12 to 24
hours.
CPR Valve
-
Transport Mode -
•
Disconnect air pipe connector and
immediately
place the
cover over it.
•
Restore the mattress by reconnecting the
supply tubes to the power unit.
To activate the CPR:
P
ull and rotate the dial of
the CPR valve to ‘click’
into the OPEN position.
At the same time, disconnect the air pipe connector
from the Pump to speed up the air release.
l
eft hand side of the mattress
To reset CPR:
Ensure the dial of the CPR
attachment is rotated until it ‘clicks’ into the
CLOSED position.