1.
Insert MD40 flange assembly
2.
Insert bolts and tighten.
Note: when tightening bolts, care must be taken not to tighten
them unevenly as this may cause the gasket not to seal properly.
STAGE 3
| Replace thimble and associated parts.
How to re-install the Drive Thimble:
1.
Push Drive Thimble over the Flange Assembly (take care when re-coupling the magnets
–
the magnetic coupling may cause a sudden movement)
2.
Replace Bearing Cap
3.
Insert and tighten 3 x M3 Grub Screws
4.
Replace Dual Shaft End Cap (ensure that the drive pin locates in the drive slot)
5.
Insert and tighten 3 x M3 Grub Screws
2.1.5 Bakeout procedure
The manual drives can be baked to 250
o
C without the need to remove any components.
In the case of the dual shaft version
, please refer to the manufacturer’s instructions relating to any
motors/pulleys you may have installed onto the MagiDrive.
End Cap
3 x M2.5 Grub Screws
Bearing Cap
Drive Thimble
3 x M2.5 Grub Screws
Содержание MD40
Страница 40: ...MD40 PRODUCT INSTRUCTION MANUAL MD40 Page 40 of 55 Author N Carapiet Revision 1 1 3 1 3 Switch Pin Out ...
Страница 42: ...MD40 PRODUCT INSTRUCTION MANUAL MD40 Page 42 of 55 Author N Carapiet Revision 1 1 3 2 3 Home Sensor Pin Out ...
Страница 46: ...MD40 PRODUCT INSTRUCTION MANUAL MD40 Page 46 of 55 Author N Carapiet Revision 1 1 Drawings Base Drive ...
Страница 47: ...MD40 PRODUCT INSTRUCTION MANUAL MD40 Page 47 of 55 Author N Carapiet Revision 1 1 Dual Shaft ...
Страница 48: ...MD40 PRODUCT INSTRUCTION MANUAL MD40 Page 48 of 55 Author N Carapiet Revision 1 1 Pneumatic Drive ...
Страница 54: ...MD40 PRODUCT INSTRUCTION MANUAL MD40 Page 54 of 55 Author N Carapiet Revision 1 1 NOTES ...