Instructions for use (rev 05/18)
e.ON SINTER
UGIN DENTAIRE
34
When the desired changes of program have been made, press "SAVE" key to save the complete
program.
After validation, the “Working interface” displayed on the screen.
6.4.2.
Programming instructions
C: Temperature in degrees Celsius (°C)
T: Time in minutes (min)
-
Temperature 1: used as the temperature start value (usually set to 0)
-
Time 1: total duration of first stage (0 to 9999 minutes).
-
Temperature 2: final temperature of first stage and start temperature of second stage.
Temperature values from C1 to C2 define the increase rate, the recommended value must be
less than 10 °C/min.
-
Time 2: total duration of second stage (0 to 9999 minutes).
-
Temperature 3: final temperature of second stage and start temperature of third stage. If C3 is
less than C2 (temperature decrease), the recommended value must be less than 30 °C/min.
-
Time 3: total duration of third stage (0 to 9999 minutes).
-
Temperature 3: final temperature of third stage and start temperature of fourth stage. If
necessary, the parameters are programmed as for the previous steps.
Last time "-121" indicates the end of program, temperature controller works according to the
temperature/time parameters setting. When the program detects a "-121", the control program
execute “STOP” command to return to the first program segment. STOP command is activated
with return to the previously selected program. "-121" also indicates correct cycle sequence.