7
PRE TEST CHECKLIST
SETTING INLET TEMPERATURE
ANALYZER CONNECTIONS
• Turn on and zero the analyzer, without the flue probe connected, to use ambient temperature
• Connect the flue probe thermocouple to T1 during zero countdown use for inlet ducted system
• Connect a thermocouple to T2 to measure second temp source
• Clean particle filter
• Water trap and probe line are empty of water
• Power on and zero
• All hose and thermocouple connections are properly secured
• Flue gas probe is sampling ambient FRESH air
• Water trap is fitted correctly to the instrument
• Flue temperature plug is connected
• Ensure water trap plug is inserted correctly
WARNING
Turning the pump off while the probe is in the flue will leave toxic gases inside the analyzer. Once
data has been printed or copied, it is advisable to purge the unit with fresh air as soon as possible.
To do this remove the probe from the flue and turn ON pump. Always allow the readings to return to
zero (20.9% for O
2
) prior to shutting the unit off. The meter will not switch off until the CO reading is
below 20 ppm.
CAUTION
The probe will be hot from flue gases. Remove the probe from the flue and allow it to cool naturally.
Do not immerse the probe in water, as this will be drawn into the analyzer and damage the flue
probe and the pump and sensors.
NOTE:
Take care when
inserting the temperature
probes as the pins are
polarized. Insert with the
smaller pin (+) to the right.
T2
T1
P2
P1
• Remove the protective rubber boot
• Slide the water trap unit from the analyzer
• Remove the particle filter from its’ spigot and replace
• Slide the water trap back into position and replace the protective rubber boot
EMPTYING & CLEANING THE IN-LINE WATER TRAP
CHANGING THE PARTICLE FILTER
QUICK START
Turn on the analyzer by pressing the On/Off Button for 2 seconds until the unit activates. As described the analyzer will perform a 60
second turn on purge. Once completed simply select the reading desired by the position of the rotary dial.
• Remove the rubber plug
• Allow the water to drain out
• Re-insert the rubber plug