The contamination of the objective lenses surface may
cause inaccurate observation of the fiber position, result-
ing in higher splice loss and poor performance of the
Swift KR��. Therefore, the two objective lenses have to be
cleaned on a regular basis. Otherwise, the dust sticks on
the surface and it may become irremovable eventually.
MAINTENANCE OF SPLICE QUALITY
REGULAR MAINTENANCE
AND CLEANING
Object Lens Cleaning
It is crucial to perform inspection and cleaning on a regular basis
to maintain the splice quality of the Swift KR��.
� Turn the power off before cleaning the object lens.
� Remove electrodes
� The lens in a circular motion from the center with an
alcohol-soaked soft cotton swab as below picture.
Remove remaining alcohol on the surface of the lenses
with dry and clean cotton swab.
� Check to ensure there is no line, scratch or stain.
� Re-assemble electrodes correctly.
� Turn on the power and perform a diagnostic test.