TYFO TYFOCOR Скачать руководство пользователя страница 1

 

 

 

 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

 

General 

 

The refractometer (Fig. 1) allows rapid and accurate 
check of freezing points of coolants and heat-transfer 
media based on ethylene glycol and propylene glycol 
by reading from the respective scales (Fig. 2). The 
refractive index (ND) of the sample fluid is simulta-
neously displayed. Battery acid can also be checked 
with the aid of this measuring device. 

 

Calibration procedure 

 

1.

 Open the illuminating plate (B), apply 1-2 drops 

of distilled water onto the prism surface (A) by using 
the pipette contained in the test set and press the 
plate onto the prism surface.

 

 

2. 

Turn the pointy end of the refractometer to the light 

and look into the eyepiece (D). Focus the scale to 
your vision by rotating the eyepiece until the scale is 
clear and easy to read. Rotate the adjusting screw 
(C) with the screwdriver included in the test set so 
that the light / dark boundary line evens up with the 
zero line (Fig. 2, WATERLINE). After the calibration 
procedure, dry the surfaces properly. 

 

Measurement procedure 

 

1.

 Open the illuminating plate, apply 1-2 drops of test 

fluid onto the prism surface, close illuminating plate.  

 

 

2.

 

Turn

 

the

 

pointy

 

end of the refractometer to the

 

light, 

rotate the eyepiece until the light / dark boundary line 
becomes visible. The line indicates the 

freezing 

point

 of TYFOCOR

®

 / water mixtures (reading from 

the ethylene glycol scale), and the 

frost protection

 

(reading from the propylene glycol scale) in case of 
TYFOCOR

®

 L / water mixtures. These terms are ex-

plained on page 2.

 

  

Instructions for measuring

 

 

1.

 Place a few drops of a sample liquid on prism sur-

face, make sure that it is entirely covered.

 

The light / 

dark boundary line cannot be seen clearly if too 
much or too little sample liquid is present.

 

After 

measuring, the sample liquid must be removed from 
the prism surface by carefully rinsing it with water 
and drying with a clean cloth. Incorrect reading can 
result from a poorly cleaned and dried prism surface.

 

 

2.

 If the refractometer is contaminated by oil, grease 

or similar matter, an accurate measurement is not 
possible, since the fluid will be repelled from the 
prism surface. In this case clean the surface with a 
spiritus soaked cloth and dry it properly afterwards. 

 

3.

 Do not damage the prism surface. Do not clean 

the refractometer under running water. It is splash-
proof, but not waterproof. 

 
 

 

 T Y F O R O P     C h e m i e     G m b H 
 ____________________________________________________________________________________ 

   

Anton-Rée-Weg 7    D-20537 Hamburg    Tel./Fax: +49 (0)40 20 94 97-0/20    e-mail: [email protected]          04/12, p.1

 

      

 Refractometer

 

  Instructions for Use 

       

  

Freezing point measurement of  

  TYFOCOR

®

 and TYFOCOR

®

 L / water mixtures 

 

  

      

 

 

 

 A         B                   C                                          D 

Fig. 2

Fig. 1 