35
Measuring Frequencies
3.3
Measuring Frequencies
The frequency can be checked. The frequency display is auto
ranging. The AC voltage range can be changed by pressing the
RANGE key.
• If signals out of the range of frequency measurement are measured,
“
−−−−−
” appears. Be aware of it.
• In a measurement environment with a large amount of noise, the
frequency may be displayed even with no input. This does not
indicate a malfunction of the instrument.
• The sensitivity of the frequency measurement is regulated by range.
(Minimum sensitivity voltage (p. 60))
When the value is less than the minimum sensitivity voltage, the
indicated value may
fl
uctuate. When the voltage range is lowered,
the value stabilizes. This does not apply to cases where the value
fl
uctuates due to noise.
• During the measurement of low frequency voltage, if the auto range
does not stabilize and the frequency cannot be measured,
fi
x the
voltage range and measure again.
The frequency is measured.
1
1
22