Page 82
2.
0000 is displayed on the screen with the cursor blinking on the right
digit.
3.
To increase or decrease the digits, press the
UP
or
DOWN
keys.
4.
Set your password, then move the cursor to
Set
by pressing the
START/STOP
key.
5.
When
Set
is blinking, press the
UP
or
DOWN
keys to return to the
program screen.
10.7
Unloading
When the cycle ended successfully (including pressing the
START/STOP
key, or any failure, after completing the sterilization
stage) message "Cycle Ended" (and the relevant failure message, if
applicable) is displayed on the screen.
•
Verify that there is no pressure in the chamber, according to the
reading on the display. Only then you may open the door.
•
Open the autoclave. (see 10.3 Opening the door)
Warning!
To avoid sever injuries from hot steam when opening the door:
•
It is strictly forbidden to lean on the autoclave.
•
It is strictly forbidden to place your hand or any part of your
body over the door.
•
Since the autoclave is defined as a bio-hazard autoclave, a failed
cycle may leave un-sterilized waste in the chamber. Therefore
failed cycle will be followed by a slow exhaust and the door will
remain locked. To enable opening the door perform a new cycle.
The door can be opened only after a successful cycle.
•
Wear heat-resistant gloves or use the tray handle to remove the load
from the autoclave
•
On completion of the cycle, the load shall be visual inspected to
ascertain that the load is dry, and that the colour of the sterilization
indicators turned to the required colour.
10.8
Stopping the process and cancelling the ERROR
message
•
It is possible to stop the program while the autoclave is operating.
Pressing the
START/STOP
key at any stage of the process stops the
operation. If the cycle was aborted before completing the sterilization
stage, it will leave the load unsterilized.