Page 76
10.
Operating Instructions
10.1
Draining and filling the water reservoir (WR only).
10.1.1
Draining
Caution!
Before starting, Make sure that the electric cord is disconnected
and there is no pressure in the autoclave.
Allow the autoclave to cool for about an hour before removing
outer covers.
The drain valve is located on the bottom of the water reservoir (See
Rear View (WR Configuration
).
The function of the drain valve is to
drain the water reservoir.
To drain the reservoir into a bucket:
1.
Open the drain valve.
2.
Fully drain the reservoir.
3.
Close the drain valve.
Note:
Drain the water from the mineral free water reservoir once a week, and
refill with fresh mineral-free water or distilled water
10.1.2
Filling
1.
Connect the electric cord to power source.
2.
Pour distilled water through a water inlet (1) near the chamber into the
reservoir using a carafe (2) so that the glass tube fills 4-5 cm below its
upper end as shown by the yellow float (3). The water quantity is
approx. 16 liters.