Page 37
Cautions!
Both PT100 temperature sensors must be inside the bottles.
For proper sterilization, fill the bottles with approximately the
same amount of liquid.
Nominal parameters default settings
•
Sterilization temperature:121
C (250
F).
•
Sterilization time: 15 minutes.
Operations sequence:
•
(PV, pre-vacuum models only)
Pulse low/Pulse high:
at one pulse it
will build vacuum down to 25 kPa.
•
Water Inlet
: Water enters the chamber
•
Heating:
The chamber is heated by actuation of electrical heaters
until the sterilization temperature is reached.
•
Sterilization:
Sterilization temperature is maintained constant during
the sterilization time.
•
Cooling:
Forced cooling to the required end temperature, chamber
pressure remains at approx. 300 kPa and the cooling valve is
opened.
•
Fast Exhaust:
the steam is exhausted out of the chamber at a fast
rate until pressure decreases to ambient pressure.
•
Drying:
N/A
•
Ending:
The system checks that End temperature and End pressure
have been reached, and then releases the door.
5.6
Program 6: Liquid B
–
Waste Cooling (C
–
cooling
models only)
Содержание LABSCI 15LV
Страница 22: ...Page 21 3 16 Overall Dimensions 3870ELVC F D mm ...
Страница 23: ...Page 22 3 17 Overall Dimensions 3870ELV PRC D mm ...
Страница 24: ...Page 23 3 18 Overall Dimensions 3870ELVC WR D Configuration mm ...
Страница 76: ...Page 75 ...
Страница 89: ...Page 88 Safety valve ring ...