Page 34
•
(PV, pre-vacuum models only)
Pulse low/Pulse high:
at one pulse it
will build vacuum down to 25 kPa.
•
Water Inlet:
Water enters the chamber
•
Heating:
The chamber is heated by actuation of electrical heaters
until the sterilization temperature is reached.
•
Sterilization:
Sterilization temperature is maintained constant during
the sterilization time.
•
Cooling:
N/A
•
Fast Exhaust:
the steam is exhausted out of the chamber at a fast
rate until pressure decreases to ambient pressure.
•
(PV, pre-vacuum models only)
Drying:
vacuum is built up and the
air pump is working.
•
Ending:
The system checks that End temperature and End pressure
have been reached, and then releases the door.
5.3
Program 3: Liquid A
For liquids when the manufacturer recommends autoclaving at
temperatures of 121
C (250
F) for 15 minutes.
Cautions!
Both PT100 temperature sensors must be inside the bottles.
For proper sterilization, fill the bottles with approximately the
same amount of liquid.
Nominal parameters default settings
•
Sterilization temperature: 121
C (250
F)
•
Sterilization time: 15 minutes