Page 52
6.
Checking and Changing Parameters
and Other Data
Bacsoft control panel allows changing parameters of the cycle and of
the system, exporting various data to, and importing from, a USB
device or to the printer, and some other options.
Cycle parameters are changeable for Custom programs only (see
Duplicate cycles), except for the Temperature sensors, Displayed
inputs, and Dry Time.
6.1
Browsing through the menus
Now you will learn how to browse through the folders. When you read
the Directories and subdirectories chapter with links to specific
menus, you will need to know how to browse through the folders
using the autoclave control panel. Below is the instruction.
Login as User (see 6.4). The Main menu screen appears. To browse
through the menus:
1.
Press the Up and Down keys to scroll through the menus.
2.
Press the Start/Stop key to enter the next screen (i.e. to get one level
down).
3.
Repeat steps 1 and 2 to enter the next screen until you get to required
screen.
Below are the example screens for the following menu:
Cycle
Parameters\ Drying\ Dry Time:
Содержание LabSci 11-V
Страница 19: ...Page 18 8 Right service door lock...
Страница 22: ...Page 21 10 Drain outlet...
Страница 34: ...Page 33 Attention This device is not a medical device and not intended for medical use...
Страница 35: ...Page 34 During the process the stages of the cycle will be displayed on the screen...
Страница 74: ...Page 73 1 Open the printer s cover door 1 by pulling it up see Fig 2 Fig 1...
Страница 79: ...Page 78...
Страница 90: ...Page 89 Attention Use protective gloves in order not to burn your hands with the hot steam Safety valve ring...