62
9.4.3
Replacing the Drain Filter
Note:
If after successive cleaning of the filter the result is not
satisfactory proceed to the replacement of the drain filter
(paragraph 9.4.3 below).
Cautions!
Before proceeding, make sure that the electric cord is
disconnected and there is no pressure in the chamber.
Use scissors to open the filter bag and not sharp blades or
pointed instrument.
Check that the new filter has not exceeded the maximum
shelf life.
Carefully un-pack the new filter and examine it for any signs
of damage.
Remove any protective packaging before inserting the filter
into place.
Carefully Insert the new filter into the housing. Do not force.
Caution!
Make sure the autoclave is not hot before cleaning it
The DRAIN filter (1) is located inside the autoclave chamber at the bottom far
end, to reach the filter open the autoclave chamber door, remove the tray (see
below). (See below).
1. Replace the old DRAIN filter, by the new DRAIN filter (P/N: CMT511-
0048).
2. Open the autoclave chamber door, remove the tray (2), (see below).
1.
Drain filter assembly
–
P.N :CMT511-0048
1
2
Содержание AMS10-230-T
Страница 2: ...2 Web site http www tuttnauerUSA com...
Страница 34: ...34 This page is left intentionally blank...
Страница 38: ...38 3 Press on the required class type 4 Enter Password 5 Click on set 6 Class setting is in progress press OK...
Страница 47: ...47 Showing the time left until the end of the process...
Страница 71: ...71...
Страница 84: ...84 8 Chamber Clean Program Manual fill clean water tank 9 Chamber Clean Program Water tank clean waste...
Страница 85: ...85 10 Select Chamber Clean Program 11 Chamber Clean Program Start Cycle...
Страница 86: ...86 12 Chamber Clean Program Water tank alarm 13 Chamber Clean Program Set manuel mode fill clean water tank...