61
9.4.2
Cleaning the Drain Filter
Caution!
Make sure the autoclave is not hot before cleaning it
Cautions!
Before proceeding, make sure that the electric cord is
disconnected and there is no pressure in the chamber.
1.
Clean the drain filter, every month.
The DRAIN filter (1) is located inside the autoclave chamber at the
bottom far end, to reach the filter open the autoclave chamber door,
remove the tray (see below).
2. Open the autoclave chamber door, remove the tray (2), (see above).
3. Clean with a soft cloth the filter and the area around the filter.
4. Check that the autoclave function normally after the drain filter cleaning.
Note:
If by cleaning the filter the result is not satisfactory, proceed to the replacement
of the drain filter (paragraph 9.4.3 below).