57
9
Preventive and Scheduled Maintenance to be
performed by the Operator
The maintenance operations described in this chapter need to be followed as
indicated to keep the device in good working condition and to keep any
breakdown time to a minimum.
The instructions that follow can easily be carried out by the operating
personnel and do not require a service technician.
Note:
Technician manual describes the maintenance operations required from
qualified technician, every two months and once a year.
Should the need arise, technical assistance or a service technician can be
requested by either calling your dealer or Tuttnauer Europe.
Daily Maintenance
Caution!
Make sure the autoclave is not hot before cleaning it.
Turn the unit on momentarily to allow the door to be opened. Open the door, unplug
the autoclave again, and proceed with cleaning.
Clean door gasket with a mild detergent, water and a soft cloth or sponge. Check
visually that the gasket is intact, not loose and clean.
Weekly Maintenance
Caution!
Make sure the autoclave is not hot before cleaning it.
Turn the unit on momentarily to allow the door to be opened. Open the door, unplug
the autoclave again, and proceed with cleaning.
Clean the outer parts of the autoclave with a soft cloth.
9.2.1
Chamber Clean
Once a week perform chamber clean process per section 12
Содержание AMS10-230-T
Страница 2: ...2 Web site http www tuttnauerUSA com...
Страница 34: ...34 This page is left intentionally blank...
Страница 38: ...38 3 Press on the required class type 4 Enter Password 5 Click on set 6 Class setting is in progress press OK...
Страница 47: ...47 Showing the time left until the end of the process...
Страница 71: ...71...
Страница 84: ...84 8 Chamber Clean Program Manual fill clean water tank 9 Chamber Clean Program Water tank clean waste...
Страница 85: ...85 10 Select Chamber Clean Program 11 Chamber Clean Program Start Cycle...
Страница 86: ...86 12 Chamber Clean Program Water tank alarm 13 Chamber Clean Program Set manuel mode fill clean water tank...