54
Opening the door and Unloading
1. Push the
button to confirm the "Successful" or the "error"
message to unlock the door.
2. Open the door.
Caution!
Open the door a little to release the steam from the chamber. Only after
the steam escaped, open the door widely.
Warning!
To avoid severe injuries from hot steam and condensed hot water that
may drip out when opening the door, it is strictly forbidden to lean on the
autoclave. It is strictly forbidden to place your hand or any part of your
body over or under the door.
3. Use the tray handle or wear heat-resistant gloves to remove the load
from the autoclave.
4. After unloading, visually inspect the load to ascertain that it is dry, and
that sterilization indicators have made the required color change.
Caution:
Water droplets and visible signs of moisture on sterile packaging or the
tape used to secure it, may compromise sterility of processed loads or be
indicative of a sterilization process failure. Visually check outside
wrapper for dryness. If there are water droplets or visible moisture on the
exterior of the package or on the tape used to secure it, the pack or
instrument tray is considered unacceptable.
Warning!
The sterility of the instruments processed in unwrapped cycles cannot be
maintained if exposed to non-sterile environment.
Содержание AMS10-230-T
Страница 2: ...2 Web site http www tuttnauerUSA com...
Страница 34: ...34 This page is left intentionally blank...
Страница 38: ...38 3 Press on the required class type 4 Enter Password 5 Click on set 6 Class setting is in progress press OK...
Страница 47: ...47 Showing the time left until the end of the process...
Страница 71: ...71...
Страница 84: ...84 8 Chamber Clean Program Manual fill clean water tank 9 Chamber Clean Program Water tank clean waste...
Страница 85: ...85 10 Select Chamber Clean Program 11 Chamber Clean Program Start Cycle...
Страница 86: ...86 12 Chamber Clean Program Water tank alarm 13 Chamber Clean Program Set manuel mode fill clean water tank...