12.2
Chamber clean Cleaning procedure
Important
The Chamber Clean cleaning program for class B and S for T-Edge10
autoclaves is described below:
The chamber must be cold.
1. All steps in this procedure must be completed without interruption.
2. Open the autoclave chamber door.
3. Remove instruments and any load, trays from the autoclave
chamber.
4. Place the trays and tray holder in the sink to be cleaned with a
stainless steel Safe cleaner.
5. Place one 6g cleaning tablet (1) from the tablet packing (2) inside
the sterilization chamber at the back bottom of the cold chamber.
(see figure below). Under the 269°F (132°C) program conditions.
Chamber clean program 6g Tablet position
6. Close the autoclave door.
7. Fill fully (Approx. 777 Ounces for T- Edge 10) clean water
reservoir with
distilled water only
. (Either Automatic or manual
filling). For manual filling, select 0, Global select 1 for automatic
filling. Fill the reservoir full until the screen indicate full.
Page 90