background image

Operation and Maintenance Manual

Cat. No. MAN205-1000000EN Rev B

Tuttnauer Europe B.V.,

Hoeksteen 11 4815 PR P.O. Box 7191 4800 GD Breda,

The Netherlands

Tel: 31 (0) 765423510,

Fax: 31 (0) 765423540

Pre-vacuum Steam Sterilizer with Generator

Model 5075

 GS-2A

, 50125 GS-2A

Содержание 50125 GS-2A

Страница 1: ...ual Cat No MAN205 1000000EN Rev B Tuttnauer Europe B V Hoeksteen 11 4815 PR P O Box 7191 4800 GD Breda The Netherlands Tel 31 0 765423510 Fax 31 0 765423540 Pre vacuum Steam Sterilizer with Generator...

Страница 2: ......

Страница 3: ...ual Pre vacuum Steam Sterilizer with Generator Cat No MAN205 1000000EN Rev B Tuttnauer Europe B V Hoeksteen 11 4815 PR P O Box 7191 4800 GD Breda The Netherlands Tel 31 0 765423510 Fax 31 0 765423540...

Страница 4: ...https tuttnauer com...

Страница 5: ...NG CLEAN SIDE VIEW 5075GS 2A 13 3 15 FRONT PANEL LOADING DIRTY SIDE VIEW 5075GS 2A 14 3 16 REAR PANEL UN LOADING CLEAN SIDE VIEW 50125GS 2A 15 3 17 FRONT PANEL LOADING DIRTY SIDE VIEW 50125GS 2A 16 4...

Страница 6: ...DUE TO CYCLE FAILURE 35 9 CHECKING AND CHANGING PARAMETERS 35 9 1 QUICK OPTIONS SCREEN TOUCH SCREEN 35 9 2 LOGGING IN 40 9 3 SYSTEM PARAMETERS 40 9 4 MAINTENANCE 41 10 MAINTENANCE INSTRUCTIONS 42 10...

Страница 7: ...ars against defects in materials and workmanship This chamber warranty went into effect January 1997 This warranty does not apply to any instrument that has been subjected to misuse neglect accident o...

Страница 8: ...a Cut off pressure switch b Cut off thermostat The main power is shut off if there is lack of water or the pressure is above the safe level Below are the safety instructions 2 2 Safety Instructions 1...

Страница 9: ...requirements ONLY NON HAZARDOUS LIQUIDS SHALL BE DISPOSED IN PUBLIC SEWAGE 3 General Information 3 1 Introduction The GS 2A autoclave is designed for sterilization of wrapped and unwrapped instrument...

Страница 10: ...sia psig or in kPa according to the operator s requirements When the pressure is displayed in psig the atmospheric pressure is shown at sea level as 0 psig If the pressure is defined in psia or kPa th...

Страница 11: ...Tray dimensions Big 47 3 x 70 x 2 cm Small 30 3 x 70 x 2 cm No of trays 2 Load No counter Counting from 0 to 9999 and nullifies 3 3 Generator s Steam Data Property 18kW Max working pressure 3 BarG Saf...

Страница 12: ...void any injury by electrical hazard it is recommended that a ground fault protection device GFCI be installed in the electrical panel feeding the autoclave local codes may make this mandatory The ele...

Страница 13: ...cs and maximum acceptable level of contaminants indicated in the table below Physical Characteristics and Maximum acceptable contaminants levels in steam for sterilizers According to EN 13060 2010 Ele...

Страница 14: ...in accordance with acknowledged analytical methods by an authorized laboratory 3 11 Directives and Standards Every autoclave meets the provisions of the following Directives and is in compliance with...

Страница 15: ...Page 11 3 12 Overall Dimensions 5075GS 2A...

Страница 16: ...Page 12 3 13 Overall Dimensions 50125GS 2A...

Страница 17: ...Page 13 3 14 Rear Panel Un Loading Clean Side View 5075GS 2A No Description No Description 1 Touch screen 4 On off switch 2 Chamber Door 5 Service door 3 Wheel 3 4 2 1 5...

Страница 18: ...re gauge 14 Power cable socket 5 Chamber door 15 Tap water inlet for drain cooling 6 Tap water reservoir Left opening overflow leave open Right opening inlet to be connected 16 Service door 7 Tap wate...

Страница 19: ...Page 15 3 16 Rear Panel Un Loading Clean Side View 50125GS 2A No Description No Description 1 Touch screen 4 On off switch 2 Chamber Door 5 Service door 3 Wheel 3 4 2 1 5...

Страница 20: ...3 Chamber pressure gauge 13 Service door 4 Generator pressure gauge 14 Mineral free water inlet to be connected 5 Chamber door 15 Mineral free water overflow leave open 6 Tap water inlet to be connect...

Страница 21: ...start stop cycles to open and close doors to enter quick options and to login The screenshot below shows the real time display fields and buttons of the Touch screen No Function No Function 1 Program...

Страница 22: ...ng the sequence of operations of the cycle The measured values of temperature and pressure are printed at fixed time intervals according to the phase of the process from the bottom up beginning with t...

Страница 23: ...ee Fig 3 4 Fig 3 7 Close the printer s cover door 3 by pressing corner 2 with the tip end of the paper emerging from the slot 1 See Fig 2 Notes on treatment of thermal papers Store the papers in a dry...

Страница 24: ...on to close door 1 This symbol displays when door 2 is open Tap S the door open button to close door 2 Door is open This message is displayed when the door is opened In stand by if start key is presse...

Страница 25: ...Wrapped 121 121 C 250 F 20 20 default Range 20 30 7 Prion 134 C 273 F 18 20 default Range 0 99 8 Vacuum test Vacuum test 9 Bowie Dick test 134 C 3 5 2 During the process the stages of the cycle will...

Страница 26: ...m purple to yellow If the vacuum test is completed the screen color will remain purple Even if the vacuum test is completed the operator shall check the test results and consider whether the test resu...

Страница 27: ...to operation the following preparing and checkout procedure for the generator is to be fulfilled Warning When servicing the autoclave disconnect all the electrical power supplies to the unit This is d...

Страница 28: ...d 5 Container systems should be scientifically proven to suit the specific sterilization program Paper paper bags are used for safe post sterilization handling Usually as single wrap Intended for smal...

Страница 29: ...ts of the instrument 25 Tilt on edge items prone to entrap air and moisture e g hollowware so that only minimal resistance to removal of air the passage of steam and condensate will be met 26 Load ite...

Страница 30: ...rying of the instruments Contact your dealer for details 4 Verify that the packaging method is in accordance with good practice approach and the packaging materials are in accordance with the applicab...

Страница 31: ...Leave space between the packs or wrapped instruments to allow complete removal of air pockets in the air removal stage and for free penetration of steam in the heating and sterilization stages 8 Oper...

Страница 32: ...ror messages 3 If any messages appear on the Touch screen take the Corrective Action recommended in 4 2 The Touch screen is used to select sterilization programs You can only select programs from the...

Страница 33: ...llustrations 7 To find the desired sterilization program search through the screens by tapping or etc Up to 8 pages are possible 8 Open the door 9 If there is a need to close the door and then re open...

Страница 34: ...eans that the autoclave is ready to run a cycle 14 Tap on the Autoclave Control screen to run the selected program A confirmation screen appears asking the user to approve starting a new cycle 15 Tap...

Страница 35: ...s to waiting for user confirmation The buzzer sounds 3 times The autoclave enters Abort Cycle sequence If the reason for the alarm no longer exists tap to reset the alarm Warning The load has not comp...

Страница 36: ...move the sterilized material from the loading cart following the steps indicated in the Un loading paragraph Warnings Loading or unloading of material from the chamber can only be done if the door is...

Страница 37: ...ng the Cycle Unloading When the cycle has ended successfully including the user pressing the START STOP key after the sterilization stage has finished the message Cycle Ended is displayed In the event...

Страница 38: ...ng traffic to avoid the possibility of touching the hot load and without air movement air conditioning etc Note Do not touch the hot load Hot loads absorb moisture and may absorb bacteria from a hand...

Страница 39: ...ce your hand or any part of your body over the door 9 At the end of each working day close the main water valve 8 8 Stopping the Process due to Cycle Failure The cycle can stop itself if the unit dete...

Страница 40: ...hed USB device A folder named Application New will be created on the USB device containing all the dll and exe files of the application Export settings Exports the gain and offset of the machine 2 fol...

Страница 41: ...ports for the last 10 cycles 50 cycles or all cycles To print cycles tap The Print cycles screen appears 1 Select the desired Print cycle option 2 Tap 3 Tap to return to Quick options menu screen 4 Ta...

Страница 42: ...e user to set up the Date and the Time on the machine The operator should perform set Date and Time when turning the machine on for the first time To set date and time tap The set date and time screen...

Страница 43: ...allowed while Start cycle by clock is enabled To Start Cycle by Clock tap The Start Cycle by Clock screen appears 1 The hour is displayed in HH MM format The hour range is 24 hours i e from 0 to 23 2...

Страница 44: ...r name Admin is the default and the password 0001 is the default then tap the Login icon You will see the main menu screen Admin version see below Note If user already logged in pressing will open the...

Страница 45: ...nance menu The available sub menus Handle Counters Printer Test Handle counters Reset Atmospheric Pressure The screen Handle counters offers the user the option to reset a counter from list of counter...

Страница 46: ...toclave with a soft cloth Clean and descale the chamber Periodically by the operator Once per month clean the water outlet strainer see 10 3 Cleaning frequency may be reduced according to experience C...

Страница 47: ...d the filter will come out 2 The filter and cover are pressed together Place two fingers between the filter and the cover and while supporting the cover pull the filter and cover apart 3 Insert a new...

Страница 48: ...oo high or too low from the normal range Wait until the the chamber reaches the normal range pressure RTC Error Please Set Current Date and Time This message is displayed in order to set the date and...

Страница 49: ...Analog Input Error Time Error Periodical check time exceeded Please call for service Cycle counter exceeded Please call for service The following messages require you to turn the autoclave off and on...

Страница 50: ...Cat No MAN205 1000000EN Rev B COPYRIGHT...

Отзывы: