T-HVA02
High-speed High-voltage Amplifier
Operating Instruction Manual
Страница 1: ...T HVA02 High speed High voltage Amplifier Operating Instruction Manual...
Страница 2: ...This operating instructions manual was created to be thorough but if anything is unclear or requires correction please contact the Service Support Section of Turtle Industry 4 We will bear no responsi...
Страница 3: ...n overheating and fire or electricity leaking Do not use the product anywhere it could be exposed to water or chemicals Failure to do so could result in fire or other type of disaster Do not use the p...
Страница 4: ...al INPUT 8 3 2 Output Terminal OUTPUT 8 3 3 Output Switch OUTPUT CONTROL 8 3 4 Offset Adjustment OFFSET 9 3 5 Output Monitor Terminal OUTPUT MONITOR 9 3 6 GND Terminal GND 9 3 7 Fuse FUSE 9 3 8 AC Inl...
Страница 5: ...se ensure you have thoroughly read and understood this document in ensuring correct use of the product 1 1 Product Outline This product is a high speed direct current amplifier with an output voltage...
Страница 6: ...justment do not exceed the range of 300 V Output Switch A push button switch used to switch the output signals on and off Lighting up red means signals are being output Even when the output has been s...
Страница 7: ...or Terminal A monitor terminal for the output terminal About 1 100 voltage of the output voltage will be output The effective bandwidth is less than 1 MHz To prevent the equipment from being damaged d...
Страница 8: ...er very dangerous Before connecting ensure sure to cut off the power supply GND Terminal Connects to the internal signal GND The internal signal GND the earth terminal for the AC inlet and the case ar...
Страница 9: ...f the human body easily coming in contact with the centerline The operating output voltage is from 300V to 300V The OUTPUT shield side and INPUT shield side have common potentiality electrically conne...
Страница 10: ...of the offset adjustment does not exceed 300 V 3 5 Output Monitor Terminal OUTPUT MONITOR The monitor terminal for the output amplifier which monitors the output terminal of the internal amplifier and...
Страница 11: ...ure The overheat protection will not recover until the power supply has been cut off Also has an overcurrent protection circuit The protection circuit will be active while the rated output current is...
Страница 12: ...sufficient and detailed proof in writing of the defect If the fault has been caused by misuse or abnormal conditions of operation repairs must be paid for Prior to repair in this instance a quotation...
Страница 13: ...l for T HVA02 12 Turtle Industry Co Ltd Support Service Section Technical Division Technical Department Email tokyo turtle ind co jp FAX 81 29 843 2024 Address 1 12 4 Nishineminami Tsuchiura shi Ibara...
Страница 14: ...m Load Resistance 2 k resistance load Input Resistance 1 M Gain 30 times 29 5 dB fixed 3 Output Offset Adjustment approx 50 V Small Signal Frequency Property DC 2 MHz 3 dB no load of 15 Vpp Power Supp...
Страница 15: ...14 6 2 Outline of Dimensions AC100V POWER INPUT OUTPUT MONITOR OFFSET 50V 50V OUTPUT CONTROL 1 100 GAIN 30 T HVA02 High Voltage Amplifier Max10Vpk 95V 105V AC FUSE TIME LAG 3A OUTPUT Max400Vpk DANGER...
Страница 16: ...Instruction Manual for T HVA02 15 Instruction Manual for T HVA02 Date of Issue April 2012 Second Edition Issued by Turtle Industry Co Ltd Edited by Turtle Industry Co Ltd c2011 Turtle Industry Co Ltd...