Maintenance
49
1. When using the cyclone, be sure that there is no impactor disk
installed. Use the standard inlet (unmarked) with gasket.
2. Attach the Dorr-Oliver
cyclone to the DPM
cyclone to create the
DPM assembly.
3. Attach the DPM
cyclone assembly
and sample tube
onto the inlet.
4. Adjust the flow rate to 1.7 L/min.
See
Chapter 3, “
for
instructions on how to set the
flow rate.
TSI 4146 flow meter (sold separately) being
used to calibrate flow setting
The SidePak
™ monitor and DPM cyclone assembly are now ready
to use. Attach the DPM cyclone assembly to the individual test
subject's clothing using the U-tube and clip provided with the DPM
cyclone assembly.