Chapter 3
32
Effect of Calibration Factors on Full Scale and Alarms
•
Note that the instrument maximum full scale mass
concentration reading is not fixed at 100 mg/m³; it changes
when the Photometric Calibration Factor is changed. The
same is true with the Response Factor and response
concentration reading.
•
For instance, if a Photometric Calibration Factor of 2.00 is
chosen, all readings are multiplied by 2.00, and the effective
maximum mass concentration reading of the instrument is then
200 mg/m³. Similarly, if the Response Factor is set at 25%, all
mass concentration readings are multiplied by .25, and the
effective maximum response concentration reading of the
instrument is 50 mg/m
3
.
•
Since the range of selectable Photometric Calibration Factors is
0.10 to 10.00, the effective maximum full scale mass
concentration reading can be anywhere between 10.0 and
1000 mg/m³.
•
The selectable range of 0% to 100% for the Response
Factor limits the maximum response concentration to
between 0.00 mg/m
3
and the current full scale mass
concentration reading.
•
Since all alarms must be allowed to approach full scale, the
maximum alarm value is 999 mg/m³. This creates a situation
where the alarm values can be set higher than the maximum
reading. In this case, the instrument will never be able to read
high enough to trigger the alarm.
•
To prevent this situation, the instrument displays the message
“
Warning: Alarm value is out of range
” if you try to enter a
value that is out of range.
Содержание SidePak AM520
Страница 8: ...viii This page intentionally left blank...
Страница 22: ...6 Chapter 1 This page intentionally left blank...
Страница 31: ...Operation 15 Battery Screws Remove these two screws to remove the battery...
Страница 56: ...Chapter 3 40 This page intentionally left blank...
Страница 78: ...Chapter 5 62 This page intentionally left blank...
Страница 88: ...Appendix B 72 This page intentionally left blank...
Страница 90: ...Appendix C 74 This page intentionally left blank...
Страница 91: ...75 Appendix D CSA Certificate of Compliance Certificates of Compliance on file at TSI...
Страница 92: ...Appendix D 76 AM520 Declaration of Conformity...
Страница 93: ...CSA Certificate of Compliance 77...
Страница 94: ...Appendix D 78 AM520i Declaration of Conformity...
Страница 95: ...CSA Certificate of Compliance 79...
Страница 96: ...Appendix D 80 This page intentionally left blank...
Страница 97: ......