PERSONAL IMPACTORS
MODEL 200
USER GUIDE
P/N 6011802, REVISION A
JUNE 2018
Страница 1: ...PERSONAL IMPACTORS MODEL 200 USER GUIDE P N 6011802 REVISION A JUNE 2018 ...
Страница 2: ......
Страница 3: ...ple of Operation Model 200 Personal Impactor Description Operation Specifications Spare Parts and Accessories List Technical Contacts Returning for Service References Appendix A Impactor Flow Calibration Instructions Appendix B Typical Model 200 Impactor Pressure Drops ...
Страница 4: ...ii Manual History The following is a manual history of the Personal Impactors Model 200 User Guide P N 6011802 Revision Date A June 2018 ...
Страница 5: ...y anyone other than a factory authorized service center with the one exception where requirements set forth in the manual allow an operator to replace consumables or perform recommended cleaning g This warranty is VOID if the product has been misused neglected subjected to accidental or intentional damage or is not properly installed maintained or cleaned according to the requirements of the manua...
Страница 6: ...h contains the complete and exclusive limited warranty of Seller This LIMITATION OF WARRANTY AND LIABILITY may not be amended modified or its terms waived except by writing signed by an Officer of Seller Service Policy Knowing that inoperative or defective instruments are as detrimental to TSI as they are to our customers our service policy is designed to give prompt attention to any problems If a...
Страница 7: ...tion 2 Operation 4 Disassembly Procedure 4 Coating the Impaction Plate 4 Assembly Procedure 5 Operating Procedure 5 Specifications 6 Spare Parts and Accessories List 6 Technical Contacts 7 Returning for Service 8 References 8 Appendix A Impactor Flow Calibration Instructions 9 Appendix B Typical Model 200 Personal Impactor Pressure Drops 10 ...
Страница 8: ...vi Personal Impactor Model 200 This page intentionally left blank ...
Страница 9: ...ilable for gravimetric and or chemical composition analysis The Personal Impactor obtains personal exposure data while the wearer goes about their normal activities An appropriate personal sampling air pump worn by the user is necessary to provide the necessary air flow See Appendix B for typical impactor pressure drops NOTE Six versions of the Model 200 Impactors are available from TSI Incorporat...
Страница 10: ...treamlines and are collected on the after filter M o d e l 2 0 0 P e r s o n a l I m p a c t o r D e s c r i p t i o n A schematic diagram of the Model 200 Personal Impactor is shown in Figure 2 The sampler consists of three basic parts a nozzle cap an impaction ring assembly and a base The impaction ring serves both as an impaction surface and as a clamping ring for the after filter The after fil...
Страница 11: ...User Guide 3 Figure 2 Schematic Diagram of Personal Impactor Phillips flat head screws are normally used instead of the spanner screws ...
Страница 12: ...action ring assembly is removed exposing the after filter Coating the Impaction Plate To keep the particles from bouncing off the impaction ring light mineral oil should be applied to this surface For most sampling applications saturating the porous metal with a light mineral oil is best As the particles strike the ring the oil will wick up through the deposit and continually provide a fresh surfa...
Страница 13: ...e same torque is applied to both screws Another way of assembling the Model 200 Personal Impactor is with the impactor Clamping Device shown in Figure 3 This device ensures that the nozzle cap is clamped parallel to the base and that the correct clamping force is applied to the after filter The handle of the clamping device is turned until the clutch in the handle slips The screws are tightened an...
Страница 14: ...ntration at a work place environment S p e c i f i c a t i o n s Sampling flow rate 2 0 L min 4 0 L min or 10 0 L min Cut point aerodynamic diameter 2 5 m or 10 m Filter diameter 37 mm Dimension L x W x H 60 x 65 x 22 mm 2 3 in x 2 5 in x 0 9 in Weight 48 g 1 7 oz S p a r e P a r t s a n d A c c e s s o r i e s L i s t All replacement parts must be factory approved Call the manufacturer for detail...
Страница 15: ... it for service visit our website at http rma tsi com or contact TSI at TSI Incorporated 500 Cardigan Road Shoreview MN 55126 USA Phone 1 800 680 1220 USA or 1 651 490 2860 E mail answers tsi com TSI GmbH Neuköllner Strasse 4 52068 Aachen GERMANY Telephone 49 241 52303 0 Fax 49 241 52303 49 E mail tsigmbh tsi com Web www tsiinc de TSI Instruments Ltd Stirling Road Cressex Business Park High Wycomb...
Страница 16: ...A shipping address Use the original packing material to return the instrument to TSI If you no longer have the original packing material seal off any ports to prevent debris from entering the instrument and ensure that the display and the connectors on the instrument front and back panels are protected This instrument is very fragile and must be packed in a manner appropriate for a precision instr...
Страница 17: ...ching a pressure gauge Measuring the pressure between the flowmeter and the impactor is only necessary if there is a large pressure drop across the flowmeter If the flowmeter has a low pressure drop such as a mass flowmeter bubble flowmeter or a laminar flow element meter the pressure at the calibration cap does not need to be measured and the pressure tap can be closed off with a rubber cap The s...
Страница 18: ...iber filter e g Gelman Type A E PEM Description Cut Point m Flow Rate L min of Nozzles Nozzle Diameter in Nozzle P in wg Filter P in wg Total P in wg PEM 2 10 10 2 0 5 0 1200 0 003 1 37 1 37 PEM 4 10 10 4 0 10 0 1200 0 003 2 73 2 73 PEM 10 10 10 10 0 10 0 1520 0 008 6 83 6 84 PEM 2 2 5 2 5 2 0 5 0 0430 0 190 1 37 1 56 PEM 4 2 5 2 5 4 0 10 0 0430 0 190 2 73 2 92 PEM 10 2 5 2 5 10 0 10 0 0670 0 201 ...