Relative Humidity and Temperature Measurements
Relative Humidity and Temperature Screen
EVM Series User Manual
78
5.
Or select the
Detail
soft key and a summary of all the measurements will appear. (Below is an
explanation of measurements.)
Measurement
values
Explanation
LEVEL
Level
-
Notation used to represent current live reading of the sensor. (This is updated once a
second.)
MIN
Minimum
- Minimum level value which occurred during a log session.
MAX
Maximum -
Maximum level value which occurred during a log session.
AVG
Average -
Arithmetic average of all level values which occurred during a log session.
STEL
Short term exposure level
- is a calculation used to indicate an exposure level (to a toxin)
over a short-term period. This value is calculated by averaging the last 15 one minute
averages. This value is invalid for the first 15 minutes and is updated every minute thereafter.
Essentially, it is used when the building occupant/worker is exposed to high concentrations of
chemical gases. The measurement will average after 15 minutes of data logging and continue
to compute after each minute of logging, using measurements from the most recent fifteen
minutes. If less than 15 minutes has elapsed, the STEL will be invalid by displaying dashes.
TWA
Time weighted average
- is an average based on an 8-hour interval. The summing all of the
level values taken and dividing by the number of samples in 8 hours (at one sample per
second the total would be 28,800). The important thing here is that the average is calculated
the same way regardless of whether you've run the instrument for 1 hour, 8 hours, 24 hours
or any other amount of time (i.e. the number of samples used to divide by is the same). This
value is updated once a second.
Table 5-3: Measurement values explained
Relative Humidity and Temperature Measurements
The relative humidity and temperature sensor will display on the left-hand side of the screen with two
measurement values. The measurement displayed in the second column is user-selectable via
Enter
.
Relative Humidity and Temperature Screen
To select a different
measurement type, press
the
Enter
key repeatedly.
Select
,
Trend
,
Detail
soft
keys (used to change the
measurement view).
Figure 5-8: RH and temperature measurement screen
Содержание EVM SERIES
Страница 1: ...ENVIRONMENTAL MONITOR EVM SERIES USER MANUAL...
Страница 2: ......
Страница 6: ......
Страница 20: ...EVM Series Models Explained EVM Series User Manual 10 This page intentionally left blank...
Страница 52: ...Setting up Parameters via the Keypad Auto run EVM Series User Manual 42 This page intentionally left blank...
Страница 100: ...File System Menu Estimated Run Time EVM Series User Manual 90 This page intentionally left blank...
Страница 118: ...EVM Reports and Printing EVM Firmware Update EVM Series User Manual 108 This page intentionally left blank...
Страница 144: ...Warranty EVM Series User Manual 134 This page intentionally left blank...