Placement and Calibration of Sensors
Calibrating CO2 AND O2
EVM Series User Manual
60
5.
Once your calibration is set, intended and actual cal values will appear on the display screen.
Press
Save
to store these as your calibration set points.
NOTE:
To redo the span calibration, press the
Span
soft key and repeat steps 2–5.
* New value
Previous calibration result
SAVE soft key
Figure 4-22: Saved Calibration screen
6.
Turn the regulator off and disconnect the tubing from the instrument.
7.
Repeatedly press
On/Off/Esc
until you return to the start screen.
8.
Press
◄
/
►
Arrows
to view measurement screens.
Calibrating CO
2
AND O
2
With the Carbon Dioxide (CO
2
) or Oxygen (O
2
) calibration, you may perform a
zero calibration and a span calibration (when the cal gas is connected) in the
field. Each of these screens will be discussed below.
The CO
2
sensor and O
2
sensor requires a calibration gas, a regulator with at
least 1 L/min flow rate and inert tubing.
A zero calibration using a Nitrogen (N
2
) gas is recommended. For a span
calibration, the range (or span) should be a sizeable fraction of the selected
full scale range of the gas.
O
2
Accuracy
The range for O
2
sensor is 0–30% with a typical span cal of 20.9%
Calibrating a zero cal and span cal for CO
2
or for 0
2
sensor
1.
Connect
one of the following gases with the regulator and inert tubing:
N
2
gas (used for zero cal.)
CO
2
calibration gas
Room air for 0
2
calibration gas (i.e., 20.9%)
2.
Attach the regulator and tubing to the gas cylinder.
3.
Place the
Cal. cup
over the
Air inlet cover
and slide the
tubing
over the black air inlet casing.
4.
From the start screen, press
▼
Arrow
to select
Calibration
and press
Enter
.
Содержание EVM SERIES
Страница 1: ...ENVIRONMENTAL MONITOR EVM SERIES USER MANUAL...
Страница 2: ......
Страница 6: ......
Страница 20: ...EVM Series Models Explained EVM Series User Manual 10 This page intentionally left blank...
Страница 52: ...Setting up Parameters via the Keypad Auto run EVM Series User Manual 42 This page intentionally left blank...
Страница 100: ...File System Menu Estimated Run Time EVM Series User Manual 90 This page intentionally left blank...
Страница 118: ...EVM Reports and Printing EVM Firmware Update EVM Series User Manual 108 This page intentionally left blank...
Страница 144: ...Warranty EVM Series User Manual 134 This page intentionally left blank...