Placement and Calibration of Sensors
Particulates Calibration (Zero Cal)
EVM Series User Manual
50
7.
Allow the level to stabilize (approximately 1–3 minutes). It will state
Waiting
(A) and
Set
soft key
will appear when it has stabilized (B).
A
B
Zero
will appear
when it has
stabilized
Current Level
reading
Set soft key will
appear when
stabilized
Figure 4-10: Particulate calibration and waiting screen
8.
Press the
Set
soft key to set the
Zero cal
point.
9.
The instrument will show a
Pause
state with the intended and actual values.
Save soft key
Figure 4-11: Particulate calibration pause screen