Placement and Calibration of Sensors
Particulates Calibration (Zero Cal)
EVM Series User Manual
48
Zero/HEPA filter
Used for particulate zero cal.
Zero filter with coupler and vent opening
Insert the Zero Filter via the coupler into the “2.5”
vent opening.
Turret
Turn the turret so “PM” appears on the display of
the EVM. The raised section will align with the
“impactor” text.
Figure 4-6: Particulate zero cal with HEPA filter
2.
Adjust the turret/impactor to
PM
(refer to “Turret” in Figure 4-6).
3.
From the
Start
screen, press
▲
/
▼
Arrows
until
Calibration
is selected and press
Enter
.
Calibration
– select to open to
the Calibration set screen
Figure 4-7: Selecting calibration screen
4.
In the
Calibration
screen, press
▲
/
▼
Arrows
and select
PM
(for particulate matter).
NOTE
: If the impactor setting is incorrect, the instrument will flash a warning sign indicating
that you need to turn the impactor to “PM” before proceeding.
Содержание EVM SERIES
Страница 1: ...ENVIRONMENTAL MONITOR EVM SERIES USER MANUAL...
Страница 2: ......
Страница 6: ......
Страница 20: ...EVM Series Models Explained EVM Series User Manual 10 This page intentionally left blank...
Страница 52: ...Setting up Parameters via the Keypad Auto run EVM Series User Manual 42 This page intentionally left blank...
Страница 100: ...File System Menu Estimated Run Time EVM Series User Manual 90 This page intentionally left blank...
Страница 118: ...EVM Reports and Printing EVM Firmware Update EVM Series User Manual 108 This page intentionally left blank...
Страница 144: ...Warranty EVM Series User Manual 134 This page intentionally left blank...