47
Chapter 4
Maintenance
Daily Maintenance Checks
Check the DustTrak
™ II/DRX aerosol monitor a minimum of once a
day to change the battery pack(s) and make sure that the instrument
is operating properly. A visual inspection of the instrument and case
can ensure that the water trap is empty, the sampling inlet is
unobstructed and the tubing is not kinked.
Cleaning the Aerosol Inlet
Under normal operating conditions, the external aerosol inlet will
remain clean and unobstructed. Under very dirty conditions; however,
the inlet may become obstructed with vegetative debris or insects. To
ensure the inside of the aerosol inlet is clean, periodically blow
compressed air through the bottom of the inlet.
C A U T I O N
DO NOT
disassemble the Environmental Enclosure
aerosol inlet for any reason. It is
NOT
designed for
field service and reassembly.
Cleaning the Respirable Aerosol Inlet
Follow the same procedure found in the DustTrak
™ II or
DustTrak
™ DRX Aerosol Monitor user manual.
When to Change the Battery
Single battery packs are designed to provide power to the
DustTrak
™ II/DRX monitor for a minimum of 34 hours, even under
cold ambient conditions. However, if using the DustTra
k™ II/DRX
monitor with External, single battery packs will provide power for a
minimum of 23 hours.
If using the Dual Battery Wiring Harness, two battery packs will
provide power to the DustTrak
™ II/DRX monitor for roughly 65 hours
or two battery packs will provide power to the DustTra
k™ II/DRX
monitor with External Pump for at least 42 hours. To ensure
uninterrupted operation, the packs should be replaced daily or within
the time period listed above.
Содержание DustTrak 8535
Страница 2: ......
Страница 17: ...Parts Identification and Unpacking 7 Item Part Number Description 801906 Router Mounting Kit ...
Страница 18: ...8 Chapter 1 This page intentionally left blank ...
Страница 40: ...30 Chapter 2 Figure 30 Attach Power Cable to Screw Terminals ...
Страница 68: ...58 Appendix A This page intentionally left blank ...
Страница 69: ......