TSI Instruments ALNOR AIRGARD 200 Скачать руководство пользователя страница 1

 

 

FUME HOOD MONITOR 

ALNOR

®

 AIRGARD

®

 200 

 

 

 

OWNER’S MANUAL 

 
 
 
 

P/N 116670080, REV 09 

SEPTEMBER 2014 

 
 

 

 

Содержание ALNOR AIRGARD 200

Страница 1: ...FUME HOOD MONITOR ALNOR AIRGARD 200 OWNER S MANUAL P N 116670080 REV 09 SEPTEMBER 2014...

Страница 2: ......

Страница 3: ...ment or which are modified by any person other than Seller The foregoing is IN LIEU OF all other warranties and is subject to the LIMITATIONS stated herein NO OTHER EXPRESS OR IMPLIED WARRANTY OF FITN...

Страница 4: ...iv This page intentionally left blank...

Страница 5: ...quired 5 Procedure 5 Simplified Installation 7 General Installation 7 Procedure 7 Calibration 8 Equipment 8 Procedure 8 Step by Step Operation 10 Alarm Acknowledgment 10 Alarm Test 10 Horn Disable fro...

Страница 6: ...alibrate the AirGard 200 monitor The entire manual should be read first before proceeding with the actual installation and calibration of the instrument 1 MOUNTING SCREWS Two screws secure the monitor...

Страница 7: ...e the relay contacts Refer to specification section to determine the relay contact ratings before making connections to the relay The relay contacts may be used to trigger a remote alarm device The de...

Страница 8: ...nstall this jumper only when the monitor is powered down Insertion or removal of this jumper while the monitor is powered up will not have any effect on the delay configuration Pole 3 of JP1 is used t...

Страница 9: ...st be observed when installing or making repairs in the vicinity of electrical equipment Tools Required 1 Electric drill 2 Electric saw or tool for cutting sheet metal 3 Drill bit size 2 5 mm 104 4 Ph...

Страница 10: ...hose on the end of the plastic sidewall tube adapter and insert the other end through the back enclosure Step 7 Plug the power adapter into the voltage mains Insert the other end of the power adapter...

Страница 11: ...to create an opening for the air output port and the power cord Procedure Step 1 Drill two mounting holes in the back of the back enclosure and in two matching places on the fascia of the fume hood t...

Страница 12: ...or equivalent 2 Small screwdriver with a 1 8 wide flat tip Procedure Step 1 Double check installation to ascertain that the AirGard 200 monitor power supply and any ancillary parts are properly insta...

Страница 13: ...elocity for this area For this example assume that the traverse resulted in an average velocity of 70 fpm Step 6 If the red indicator is illuminated slowly turn the adjustment screw counterclockwise u...

Страница 14: ...e tested by pressing the TEST RESET button While the button is pressed the indicator and horn will be activated Horn Disable from keypad Under any operating condition the horn may be disabled by press...

Страница 15: ...4 may be closed Pole 3 of JP1 may have shorting jumper unintentionally installed If permanent horn silence is not desired disconnect power remove pole 3 shorting jumper and reconnect power WRONG ALARM...

Страница 16: ...all service requests If any malfunction is discovered please contact your nearest sales office or representative or call TSI s Customer Service department at 800 874 2811 U S and Canada only Ship to...

Страница 17: ...10 cm Alarm Setpoint Set point 80 fpm recommended Operating Temperature 55 F to 86 F 13 C to 30 C 5 to 95 RH non condensing Storage Temperature 40 F to 150 F 40 C to 65 C 5 to 95 RH non condensing No...

Страница 18: ...rporated Visit our website www alnor com for more information USA Tel 1 800 874 2811 India Tel 91 80 67877200 China Tel 86 10 8219 7688 Singapore Tel 65 6595 6388 P N 116670080 Rev 09 2014 TSI Incorpo...