This chapter describes how to operate the Model 3082 Electrostatic
Classifier using the touch-screen display. Touch the screen with your
fingers or the supplied stylus to make selections. For instructions on how to
operate the classifier using Aerosol Instrument Manager software, refer to
the
Aerosol Instrument Manager
®
Software for SMPS™
Instruction Manual
(TSI P/N 1930038).
O p e r a t i n g P r e c a u t i o n s
Before applying power, review the operating specifications for the
Electrostatic Classifier described in
When using the classifier, follow these operating precautions:
High-voltage is accessible within this instrument. Unplug the power
source
before
removing the cover to perform maintenance procedures.
The Model 3077/3077A Neutralizer contains radioactive material that is
subject to the regulations of the U.S. Nuclear Regulatory Commission
as well as local regulations. Carefully read the Model 3077 Neutralizer
operating manual to determine your legal responsibilities when using
this instrument.
The Model 3088 Advanced Aerosol Neutralizer contains an x-ray
source, which may be subject to local regulations. During normal
operation, you will not be exposed to x-ray radiation.
Do not
make any
changes, such as detaching the x-ray source from the neutralizer tube,
to the Aerosol Neutralizer.
Install the Neutralizer as described in
Setting Up the Electrostatic Classifier
If you operate the classifier without a Model 3077/3077A or 3088
Neutralizer (such as for electrospray applications), do not use the
aerosol inlet. If you do, you are not completing the aerosol flow path.
Instead, remove the flag assembly and connect the aerosol source
directly to the DMA polydisperse inlet.
Do not
operate the classifier without the impaction plate installed. If
you do, the penetration curve will be incorrect and the aerosol flow rate
cannot be measured. A “Low aerosol flow” warning may indicate the
absence of an impaction plate.
Never
use the DMA as a handle when lifting the classifier.
Содержание 3082
Страница 2: ......
Страница 32: ...1 12 Electrostatic Classifier Model 3082 and SMPS Spectrometer Model 3938 This page intentionally left blank ...
Страница 76: ...3 6 Electrostatic Classifier Model 3082 and SMPS Spectrometer Model 3938 This page intentionally left blank ...
Страница 142: ...5 54 Electrostatic Classifier Model 3082 and SMPS Spectrometer Model 3938 This page intentionally left blank ...
Страница 175: ...Maintenance Service and Troubleshooting 6 33 This page intentionally left blank ...
Страница 176: ......
Страница 189: ...Theory of Operation B 7 Figure B 4 Flow Schematic for the Electrostatic Classifier with Nano or 1nm DMA ...
Страница 222: ......
Страница 224: ......