6
Chapter 3
Menus
DISPLAY SETUP
Display setup menu is where you will setup the desired parameters to be
displayed on the running screen. With a parameter highlighted you can then
use the
ON
soft key to have it show up on the running screen or select the
OFF
soft key to turn off the parameter. Use PRIMARY soft key to have a
parameter show up on the running screen in a larger display. Only one
parameter can be selected as a primary, and up to 2 secondary parameters
can be selected at one time.
SETTINGS
Settings menu is where you can set the general settings. These include
Language, Beeper, Select Units, Time Constant, Contrast, Set Time, Set
Date, Time Format, Date Format, Number Format, Backlight and Auto Off.
Use the
g
or
h
soft keys to scroll through the settings for each option and
use the
key to accept settings.
FLOW SET UP
In
Flow Setup
mode, there are 4 types: Round Duct, Rectangle Duct, Duct
Area, and Horn. Use the
g
or
h
soft keys to scroll through the types and
then press the
key to accept the desired type. To change the value,
highlight the
Enter Settings
option and press the
key. Use the instructions
below to change the value of the type that you selected.
NOTE:
The horn numbers are the models of the horns. For example, 100
refers to a horn model number AM 100. Only horns with Model
numbers as follows can be used with this function: AM 100,
AM 300, AM 600 and AM 1200. If a horn model number is chosen,
the instrument will return to measuring mode and use a
preprogrammed curve to calculate flow rate from velocity.
ACTUAL/STANDARD SETUP
Choose Actual/Standard measurements and parameters in the Act/Std Setup
menu. Within this menu, the user can also select Standard Temperature,
Standard Pressure and a source for the actual temperature. The actual
barometric pressure must be entered to convert air velocity and volume
measurements to actual conditions.