Chapter 4
Maintenance
The Alnor CF920/A
IRFLOW
IAQ920 requires very little maintenance to keep
it performing well.
Recalibration
To maintain a high degree of accuracy in your measurements, we
recommend that you return your Alnor CF920/A
IRFLOW
IAQ920 to TSI for
annual recalibration. Please contact one of TSI’s offices or your local
distributor to make service arrangements and to receive a Return Material
Authorization (RMA) number. To fill out an online RMA form, visit TSI’s
website at
U.S. & International
UK
TSI Incorporated
TSI Instruments Ltd
.
Alnor Products
Stirling Road
500 Cardigan Road
Cressex Business Park
Shoreview MN 55126-3996
High Wycombe
Tel: (800) 874-2811
Bucks
(651) 490-2811
HP12 3RT United Kingdom
Fax: (651) 490-3824
Tel: +44 (0) 149 4 459200
Fax: +44 (0) 149 4 459700
The Alnor CF920/A
IRFLOW
IAQ920 can also be recalibrated in the field
using the CALIBRATION menu. These field adjustments are intended to
make minor changes in calibration to match a user's calibration standards.
The field adjustment is NOT intended as a complete calibration capability.
For complete, multiple-point calibration and certification, the instrument
must be returned to the factory.
Cases
If the instrument case or storage case needs cleaning, wipe it off with a soft
cloth and isopropyl alcohol or a mild detergent. Never immerse the Alnor
CF920/A
IRFLOW
IAQ920. If the enclosure of the Alnor CF920/A
IRFLOW
IAQ920 or the AC adapter becomes broken, it must be replaced immediately
to prevent access to hazardous voltage.
Storage
Remove the batteries when storing the unit for more than one month to
prevent damage due to battery leakage.