41
Chapter 4
Maintenance
You can perform routine maintenance on the P-T
RAK
®
particle counter using
the instructions below. However, you are not able to perform field
calibration of the instrument. For this reason, TSI recommends that you
return your P-T
RAK
®
particle counter to the factory for annual cleaning and
calibration. For a reasonable fee, we will quickly clean and calibrate the unit
and return it to you in “as new” working condition, along with a Certificate
of Calibration. This “annual checkup” helps ensure that the P-T
RAK
®
particle
counter is always in good operating condition.
Note:
There are no user-serviceable parts inside this instrument.
Opening the instrument case may void the warranty. TSI
recommends that you return the P-T
RAK
®
particle counter to
the factory for any required maintenance or service not
described in this manual.
Maintenance Schedule
Your Model 8525 P-T
RAK
®
Ultrafine Particle Counter needs very little
required maintenance. The few items that are suggested, however, should be
done regularly, to ensure reliable operation. Table 4–1 lists the factory
recommended maintenance schedule.
Some maintenance items are suggested each time the P-T
RAK
®
particle
counter is used. Other items are performed based upon usage level, calendar
or done as needed.
Table 4–1. Recommended Maintenance Schedule
Item Frequency
Daily zero check
Daily (or before each use)
Recharge the alcohol wick
Daily (or before each use)
Change the alcohol wick
As needed (see Chapter 5,
“
”)
Clean inlet screen assembly
Monthly, or as needed
Return to factory for cleaning and calibration
Annually
Daily Zero Check
The Daily Zero Check ensures that the instrument and alcohol cartridge are
properly assembled and free from leaks. Please see the beginning of