Maintenance 5–13
Inlet Fitting
Hex Nut
Capillary Sleeve
High Voltage
Fitting
Hex Nut
High Voltage
Shield
Capillary
Figure 5-5
Removing or Installing the Capillary
3.
Loosen the hex nut on the inlet fitting but do not remove it.
4.
Carefully pull the capillary from the capillary guide leaving the
guide behind.
!
C a u t i o n
When removing the capillary from the inlet fitting, the capillary can
become damaged if not removed carefully. Read the note below for hints
on removing the capillary without damage.
Note
: When removing the capillary slowly, the capillary tip will have
a tendency to hit the edge of the hex nut on the inlet fitting if it
is not pulled straight out from the inlet fitting. To prevent
damage to the capillary, use a steady hand and pull the
capillary straight out from the inlet fitting or pull the capillary
out slowly until it is almost removed and then pull quickly to
completely remove the capillary. One alternative solution is to
remove the inlet fitting before removing the capillary. Another
is to perform steps 5–7 before steps 3 and 4.
5.
Remove the high-voltage shield by loosening the mounting
screw and pulling the high-voltage shield upward until it clears
the capillary.
6.
Loosen the hex nut on the high-voltage fitting but do not
remove it.