4–6
Model 3480 Electrospray Aerosol Generator
Pressure Chamber
Bottom Half
Sample Vial
Figure 4-1
Inserting a Sample Vial into the Pressure Chamber
5.
Turn the pressure regulator knob clockwise until the pressure
gauge reads 3.7 psi.
6.
Turn the air rotameter knob counterclockwise to reach
1.0 L/min.
7.
Turn the CO
2
rotameter knob counterclockwise to reach
0.1 L/min.
8.
Turn the voltage adjustment knob clockwise to reach 2.30 kV.
9.
Within two minutes the liquid should be exiting the capillary.
See Chapter 6, “
” if the liquid has not made it
through the capillary after two minutes.
G e n e r a t i n g A e r o s o l s
The Electrospray operating parameters set in the “
” section should result in a stable generation of aerosol
in the Cone-Jet mode of the Electrospray (see below for an
explanation of the “Cone-Jet” mode). Once the liquid is flowing from
the capillary tip, the high voltage can be adjusted to achieve
different Electrospray operating modes. Four distinct modes (see
Figure 4-2) can be observed through the viewing window or by the
current LED display as the electric field is increased by increasing
the high voltage: Dripping mode, Pulsating mode, Cone-Jet mode,