CaloSys Modules
65
11.9.1.1. 1x Air Drying Unit
Fig. 36 1x Air Drying Unit w/ separating bottle
1
Separating bottle
2
Status LED (Cooling mode)
3
Sample IN (connection to Sample Switch Unit)
4
Temperature display
5
Status LED (Operating mode, indicates whether Control Unit is connected to mains supply)
6
Sample OUT (connection to Calibration Control Unit)
7
Collection vessel
Removal of condensate
The condensed water of the collecting vessel of the 1x Air Drying Unit and the separating bottle must
be discarded when they are half-full of condensate at the LATEST, in order to keep the amount of
moisture entering the tubing connections as low as possible.
NOTICE
– Reduced lifetime!
If Calorimetry Process Control Units and Air Pumps are powered ON while there is a leakage
in the sample line, this may reduce lifetime of devices.
•
After you have discarded condensed water, carefully reconnect collection vessel to 1x Air
Drying Unit and reconnect separating bottle to tubing.
•
Select "Status/Calo" and check for proper "SampleFlow” on "Calo Current Values" dialog (see
Software Operating Instructions – Calorimetry, Section "Status Menu, Calo").
7
1
3
6
5
4
2