CaloSys Modules
53
connector without using a pressure reducing valve set to <2bar (<29psi) above atmospheric
pressure ("Gas Inlet1...3").
This warning is given on the instrument, attached onto the rear panel.
NOTICE
- Instrument damage!
Too high operating pressure will damage the internal sensor of the Calorimetry Control Unit
and the air pump membrane.
Never connect a gas cylinder directly to an inlet connector without using a pressure reducing valve
set to <2bar (<29psi) above atmospheric pressure ("Gas Inlet1...3").
•
For CaloSys modules that are operated with the sensors of the High-Speed Sensor Units,
connect calibration gas cylinder via a pressure reducing valve set to <5bar (<72psi) above
atmospheric pressure ("Gas Inlet1") and to <2bar (<29psi) above atmospheric pressure ("Gas
Inlets2...3").
•
WARNING!
Too high operating pressure could endanger persons in the vicinity (e.g. by
tubing or other materials bursting, etc.).
Never connect a gas cylinder directly to an inlet
connector without using a pressure reducing valve set to <5bar (<72psi) above atmospheric
pressure ("Gas Inlet1") and <2bar (<29psi) above atmospheric pressure ("Gas Inlet2, 3"),
respectively.
This warning is given on the instrument, attached onto the rear panel.
NOTICE
- Instrument damage!
Too high operating pressure will damage the internal sensor of the High-Speed Sensor Unit
and the air pump membrane.
Never connect a gas cylinder directly to an inlet connector without using a pressure reducing valve
set to <5bar (<72psi) above atmospheric pressure ("Gas Inlet1") and <2bar (<29psi) above
atmospheric pressure ("Gas Inlet2, 3"), respectively.
Connections Calibration Slave Process Control Unit
11.3.2.
The calibration gas supply and the sample air supply are connected to this Control Unit. "Unit" and
"Gas" connect to the Calibration Master Control Unit. "Sample IN" connects to the sample switch
pump via Air Drying Unit.
"Sample OUT" connects to the "Sample IN" of the Calorimetry Control Unit.
Fig. 27 Calibration Slave Control Unit
1
Tube connection to Calibration Master Control Unit.
2
3-pin connection to Calibration Master Control Unit.
1
2
3
4
Содержание Calorimetry PhenoMaster
Страница 6: ...6 PhenoMaster Hardware Operating Instructions info TSE Systems com www TSE Systems com ...
Страница 13: ...CaloSys Modules 13 Confirm default installation path by pressing Next info TSE Systems com www TSE Systems com ...
Страница 15: ...CaloSys Modules 15 Press Finish info TSE Systems com www TSE Systems com ...
Страница 69: ...CaloSys Modules 69 info TSE Systems com www TSE Systems com ...