CaloSys Modules
27
Fig. 10 Disassembling discus filter (2)
1
Discus filter
2
Inner blue connector part
Used filters…
…should be disposed of in accordance with your local waste disposal regulations.
•
Select new disk filter.
•
Plug-in both disk filter ends to inner two blue connector parts.
•
After connecting, the system including the air pump can be started up again.
If you have questions or need assistance, please contact
10.6. Air Drying Unit
The silicone tubing and the glass collection vessels should first be removed from the Air Drying Units.
The Air Drying Units should be wiped off, if required, with a moistened soft, lint-free cloth or tissue to
remove any adhering dirt. A pH-neutral detergent solution can also be used for moistening. Detergent
residues should then be wiped off.
After cleaning, the Air Drying Units can be decontaminated with H2O2 vapor
before it is brought into
the test laboratory.
The glass collection vessels can be autoclaved.
New silicone tubing can be decontaminated with hot air up to 180°C (356°F).
Used silicone tubing should be disposed of according to the national safety and environmental
regulations.
10.7. Tubing
Tubing that has not come into contact with breathed-out air from animals can be decontaminated by
using H2O2. The working life of such tubing material depends on the material itself, so that all tubing
treated with H2O2 should be checked for damage at regular intervals.
In order to eliminate impairment of the animals from exposure to H2O2, the H2O2 should be allowed
to desorb for >=2.5 days, e.g., over a weekend.
2
1
Содержание Calorimetry PhenoMaster
Страница 6: ...6 PhenoMaster Hardware Operating Instructions info TSE Systems com www TSE Systems com ...
Страница 13: ...CaloSys Modules 13 Confirm default installation path by pressing Next info TSE Systems com www TSE Systems com ...
Страница 15: ...CaloSys Modules 15 Press Finish info TSE Systems com www TSE Systems com ...
Страница 69: ...CaloSys Modules 69 info TSE Systems com www TSE Systems com ...