TruSteel AutoVap30 User Manual | | v. 1.9
33 / 39
Eyes
: Immediately flush eyes with plenty of water for at least 15 minutes, occasionally lifting the
upper and lower eyelids. Get medical aid.
Skin
: Get medical aid. Wash clothing before reuse. Flush skin with plenty of soap and water.
Ingestion
: Do not induce vomiting. If victim is conscious and alert, give 2-4 cupfuls of milk or
water. Never give anything by mouth to an unconscious person. Get medical aid.
Inhalation
: Remove from exposure and move to fresh air immediately. If not breathing, give
artificial respiration. If breathing is difficult, give oxygen. Get medical aid.
General Information
: Containers can build up pressure if exposed to heat and/or fire.
As in any fire, wear a self-contained breathing apparatus in pressure-demand, MSHA/NIOSH
(approved or equivalent), and full protective gear. Vapors may form an explosive mixture with air.
Vapors can travel to a source of ignition and flash back. Will burn if involved in a fire. Flammable
Liquid. Can release vapors that form explosive mixtures at temperatures above the flashpoint.
Use water spray to keep fire-exposed containers cool. Containers may explode in the heat of a
fire.
Extinguishing Media
: For small fires, use dry chemical, carbon dioxide, water spray or alcohol-
resistant foam. For large fires, use water spray, fog, or alcohol-resistant foam. Use water spray
to cool fire-exposed containers. Water may be ineffective. Do NOT use straight streams of
water.
·
Auto-ignition Temperature: 362.8 deg C
·
(alcohol Flash Point: 13.9 deg C ( 57.02 deg F)
·
Explosion Limits: Lower: 3.3 (alcohol)
·
Explosion Limits: Upper: 18.0 (alcohol)
·
NFPA Rating: health: 2; flammability: 3; instability: 0
TROUBLESHOOTING
Checking for Vacuum Leaks
After setting up the AutoVap
™ system, turn on the vacuum pump, open the vacuum valve V3 and
close the gas ballast valve V4. The system should hold a consistent vacuum of 2-4 PSIA. If not
check all clamp connections and tighten accordingly.
Checking for Fluid Leaks
Inspect all fluid/water cooling/heating connections to ensure no leaks are present in the system.
If Swagelok fittings are leaking, tighten with the appropriate wrench. If threaded connections are
leaking, remove connections, clean both male and female sides and remove all tape.
Apply PTFE tape, re-insert and tighten. (Making sure that no tape gets into the system).
Inconsistent Spray Pattern
Check filter on inlet side of feed pump.
Remove spray tip from evaporator and check for debris.
Содержание AutoVap30
Страница 1: ...Copyright 2020 by TruSteel LLC All Rights Reserved TruSteel AutoVap30 User Manual Version 1 9 ...
Страница 3: ...TruSteel AutoVap30 User Manual v 1 9 3 39 STEP 8 SHUT DOWN 39 ...
Страница 5: ...TruSteel AutoVap30 User Manual v 1 9 5 39 ...
Страница 6: ...TruSteel AutoVap30 User Manual v 1 9 6 39 ...
Страница 7: ...TruSteel AutoVap30 User Manual v 1 9 7 39 ...
Страница 10: ...TruSteel AutoVap30 User Manual v 1 9 10 39 Pull the Red EMERGENCY button OUT Turn all H O A switches to AUTO ...
Страница 36: ...TruSteel AutoVap30 User Manual v 1 9 36 39 AV30 FLOW IMAGE CUTAWAY CLICK HERE FOR LARGER VIEW ...
Страница 37: ...TruSteel AutoVap30 User Manual v 1 9 37 39 ...