In-line turbidity transmitter
TRUBOMAT
®
GAB 1000
12-03-2015
446 M1 02 G
MES
446-02
/1
INSTRUCTIONS MANUAL
From 0.1 to 1000 FNU
Страница 1: ...In line turbidity transmitter TRUBOMAT GAB 1000 12 03 2015 446 M1 02 G MES 446 02 1 INSTRUCTIONS MANUAL TRUBOMAT GAB 1000 In line turbidity transmitter From 0 1 to 1000 FNU ...
Страница 2: ...r controlled analysis electronics and two emitter modules The sensor set generates a 4 to 20 mA output signal in accordance with the following diagram Measuring ranges are selected with the help of the DIP switches on the receiver module Page 2 446 M1 01 G 12 03 2015 Note The tubidity measuring instruments are calibred with the internationally specified standard suspension Formazin The indication ...
Страница 3: ...e signal depends on size and quantity of water drops So it s important to install the turbidity measurement system always in such a way that the pipe is filled at any time ver tically or with a siphon That s valid not only for TURBICUBE but for all systems on the market 5 WIRING 25 30 Maximal load Ohms Voltage V 1200 1000 800 600 400 200 0 10 15 20 25 30 P Not permissible range Permissible range 6...
Страница 4: ...eassemble after cleaning by following the above instructions in reverse Note Inspect the O ring before reassembly and replace it if necessary tighten the compression disc with the included tool use the rod of the suction cup as a lever Before reinstalling the transmitters and the receiver refill the fixture with liquid medium and inspect for possible leaks Observe the positioning of hole and pin N...