6–
Figure 6–14. Gauge Positioned Over Designated Test Site
Press
〈
OK
〉
to begin the measurement. The PDA displays
Special
Count in Progress
as well as the time remaining (in seconds).
Upon completion of the count, the PDA collects the counts, stores
the partial calibration, and displays
Partial calibration saved.
Press ‘New’ again to enter true density value.
Press
〈
ok
〉
to
return to the
Special Calibration Menu
.
ENTERING TRUE DENSITY.
To enter the true density after a
partial offset has been stored, press the
〈
New
〉
button on the
Special Calibration Menu
. The PDA displays
A partial
special calibration exists. Do you wish to enter the true
density value?
To enter the true density, press
〈
Yes
〉
. The PDA requests the
true density (see Figure 6–15). Enter the true density and press
〈
Ok
〉
. The PDA requests a file name for the special calibration
(see Figure 6–16). Enter a file name (up to 20 characters) and
press
〈
Ok
〉
. The name should be as descriptive as possible, to
ensure that the special calibration value can be easily identified.
The PDA displays
Special calibration saved. Do you
wish to activate this special calibration?
Press
〈
Yes
〉
to
activate the special calibration, or
〈
No
〉
to continue. The PDA
returns to the
Special Calibration Menu
.
To erase the existing partial calibration and start a new one,
press
〈
No
〉
. The PDA asks
Do you want to erase the
current partial calibration?
Press
〈
Yes
〉
to erase the stored
partial calibration and begin a new density measurement as
described above, or
〈
No
〉
to return to the
Special Calibration
Menu
.
Содержание Enhanced RoadReader Plus 3451
Страница 7: ...Model 3451 vii NOTES...
Страница 17: ...xvi NOTES...
Страница 29: ...1 12 NOTES...
Страница 33: ...2 4 Figure 2 1 Effect of Moisture on Depth of Measurement...
Страница 85: ...4 22 NOTES...
Страница 106: ...Model 3451 6 3 Figure 6 1 Offset Menu 6 OFFSETS CALIBRATION...
Страница 114: ...Model 3451 6 11 Figure 6 10 Moisture Offset Erase Prompt 6 OFFSETS CALIBRATION...
Страница 127: ...6 Figure 6 25 Enter File Name Figure 6 23 Input True Density Figure 6 24 Thin Layer Special Calibration Results...
Страница 131: ...6 NOTES...
Страница 141: ...7 10 Figure 7 5 Project Data Spreadsheet...
Страница 145: ...7 14 Figure 7 8 Project Setup Display Figure 7 9 Standard Counts Display...
Страница 147: ...7 16 Figure 7 10 Project Status Display Figure 7 11 Source Decay Display...
Страница 149: ...7 18 Figure 7 12 Stat Test Display Figure 7 13 Drift Test Display...
Страница 153: ...7 22 NOTES...
Страница 166: ...Model 3451 Appendix A 13 Figure A 3 Cleaning the Sliding and Fixed Blocks...
Страница 174: ...Model 3451 Appendix A 21 NOTES...
Страница 175: ......
Страница 183: ...Appendix B 8 NOTES...
Страница 187: ...Figure C 5 Review Stat Test Appendix C 4 Figure C 3 View Stat Counts Figure C 4 Stat Test Results...
Страница 221: ...Appendix D 6 NOTES...
Страница 223: ...Appendix E 2 STANDARD COUNT LOG Gauge Serial Number Date Moisture MS Density DS1 Density DS2...
Страница 225: ...Appendix E 4 STANDARD COUNT LOG Gauge Serial Number Date Moisture MS Density DS1 Density DS2...
Страница 227: ...Appendix E 6 STANDARD COUNT LOG Gauge Serial Number Date Moisture MS Density DS1 Density DS2...
Страница 229: ...Appendix E 8 STANDARD COUNT LOG Gauge Serial Number Date Moisture MS Density DS1 Density DS2...
Страница 235: ...Appendix G 4 NOTES...
Страница 246: ...Model 3451 Warranty 1 NOTES WARRANTY...