![Troxler 3430 Plus Скачать руководство пользователя страница 27](http://html1.mh-extra.com/html/troxler/3430-plus/3430-plus_manual-of-operation-and-instruction_1172534027.webp)
Model 3430 Plus & 3440 Plus
2–3
MOISTURE
The gauge uses the principle of neutron
thermalization
to monitor
the moisture content of a material. The gauge includes an
americium-241:beryllium (Am-241:Be) source that is fixed in the
gauge’s base. Fast neutrons emitted by the Am-241:Be source pass
into the test material. Multiple collisions between the fast neutrons
and a similarly sized mass (such as the nuclei of hydrogen atoms)
cause the neutrons to slow to the point where further collisions with
hydrogen or other materials will not continue to reduce the neutron
energy further. These neutrons are said to have been
thermalized
.
The gauge contains a helium-3 detector that is sensitive only to
thermalized, or “slow,” neutrons. As a result, the moisture counts
relate directly to the amount of hydrogen in the material. Using the
gauge calibration, the gauge software converts the helium-3 detector
counts to a moisture content. (Note that the helium-3 detector is in
the same plane as the Am-241:Be source. Therefore, moisture
measurement is similar to a density measurement taken in
backscatter mode, as described earlier in this chapter.)
The
depth of measurement
is defined in terms of a maximum depth
beneath the surface of the material being measured. Of the neutrons
counted by the gauge, 98 percent will penetrate no deeper than the
depth of measurement. The depth of measurement is a function of
moisture content. The following equations can be used to determine
the approximate depth of measurement.
Depth (inches) = 11 – (0.17
×
M), where: M = moisture in pcf
or
Depth (mm) = 280 – (0.27
×
M), where: M = moisture in kg/m
3
The normalized curve set shown in Figure 2–1 illustrates the effects
of moisture content on the depth of measurement.
2. T
HEO
RY
OF
OP
ER
AT
ION
Содержание 3430 Plus
Страница 3: ...Model 3430 Plus 3440 Plus iii ...
Страница 7: ...Model 3430 Plus 3440 Plus vii NOTES ...
Страница 10: ......
Страница 14: ...Model 3430 Plus 3440 Plus xii NOTES ...
Страница 16: ...xiv NOTES ...
Страница 30: ...2 6 NOTES ...
Страница 64: ...4 NOTES ...
Страница 98: ...7 NOTES ...
Страница 122: ...9 NOTES ...
Страница 142: ...Appendix A 20 NOTES ...
Страница 167: ...Model 3430 Plus 3440 Plus Appendix C 21 Figure C 4 Base Assembly Part Number 110015_XXXX C TROUBLESHOOTING ...
Страница 170: ...Appendix C 24 Figure C 5 Scaler Assembly Part Numbers 110791 3430 Plus and 110876 3440 Plus Continued ...
Страница 178: ...Appendix D 2 STANDARD COUNT LOG Gauge Serial Number Date MS DS Date MS DS ...
Страница 180: ...Appendix D 4 STANDARD COUNT LOG Gauge Serial Number Date MS DS Date MS DS ...
Страница 190: ...Appendix F 8 NOTES ...
Страница 194: ...Appendix G 4 NOTES ...
Страница 200: ...Appendix H 6 NOTES ...
Страница 215: ...Model 3430 Plus 3440 Plus Index 11 W Wide Area Augmentation System WAAS 2 Y YES key 3 4 INDEX ...
Страница 216: ...Index 12 NOTES ...
Страница 217: ...Model 3430 Plus 3440 Plus Warranty 1 NOTES WARRANTY ...