4-4
MOISTURE OFFSET
Some soils contain hydrogen sources other than water and/or
neutron absorbers. Since the 3430 gauge measures moisture by
determining the hydrogen content of the material and relating this to
the water content, both types of material could cause gauge readings
that differ from the true moisture. If measuring such materials, use a
moisture offset to adjust the readings.
The offset factor (
k
) is determined by comparing the moisture
content of a laboratory sample with the moisture content determined
by a gauge reading. To determine the offset factor, use the following
procedure:
✓
Take a gauge reading at the site. Record the reading (
%M
GAUGE
).
✓
Remove a sample from the measurement site, then use
laboratory methods methods (for example, oven dry, and so on)
to determine the moisture content of the sample (
%M
LAB
).
Multiple samples and measurements may be taken. Calculate
the average moisture of the samples. This average value should
be used for the offset factor calculation.
✓
Calculate the offset factor (
k
).
k =
%M
LAB
– %M
GAUGE
100 + %M
GAUGE
×
1000
NOTE
If the k value is negative, a minus sign (–) must be
entered by pressing the down arrow before entering the
first digit.
To access the
Special
functions, press
〈
SPECIAL
〉
.
Содержание 3430-M
Страница 9: ......
Страница 14: ...xii NOTES...
Страница 19: ...Model 3430 1 5 Figure 1 1 Model 3430 Gauge and Accessories INTRODUCTION...
Страница 22: ...1 8 NOTES...
Страница 25: ...Model 3430 2 3 Figure 2 1 Direct Transmission Geometry Figure 2 2 Backscatter Geometry OPERATION THEORY...
Страница 26: ...2 4 Figure 2 3 Backscatter Surface Density Effects Top Layer Effect Curves...
Страница 28: ...2 6 Figure 2 4 Effect of Moisture on Depth of Measurement...
Страница 48: ...3 20 NOTES...
Страница 62: ...4 14 NOTES...
Страница 66: ...5 20 D xxxx M xxxx Use keys Moist R xxxx to view data Moist Avg xxxx to view data...
Страница 88: ...A 14 NOTES...
Страница 111: ...Model 3430 C 17 Figure C 2 3430 Gauge Assembly MAINTENANCE...
Страница 113: ...Model 3430 C 19 Figure C 3 3430 Gauge Base Mechanical Assembly Figure C 4 3430 Source Rod Handle Assembly MAINTENANCE...
Страница 115: ...Model 3430 C 21 Figure C 5 3430 Gauge Preamplifier Assembly MAINTENANCE...
Страница 117: ...Model 3430 C 23 Figure C 6 3430 Gauge Scaler Assembly MAINTENANCE...
Страница 128: ...E 2 Date MS DS Date MS DS STANDARD COUNT LOG Gauge Serial Number...
Страница 129: ...Model 3430 E 3 Date MS DS Date MS DS STANDARD COUNT LOG Gauge Serial Number STANDARD COUNT LOG...
Страница 130: ...E 4 Date MS DS Date MS DS STANDARD COUNT LOG Gauge Serial Number...
Страница 140: ...Index 8 NOTES...
Страница 141: ...Model 3430 Warranty 1 NOTES WARRANTY...