Model 3430
3-15
TAKING A MEASUREMENT – SOIL MODE
The
Soil
mode is automatically selected when a Proctor value is
enabled (see page 3-8).
NOTE
When not taking measurements, always keep the
source rod in the SAFE position. For added operator
safety, the source rod on the 3430 gauge automatically
retracts to the SAFE position when the gauge is lifted
by the handle.
If you do not hear a
click
when the source rod is raised to the
SAFE
position, look at the bottom of the gauge to verify that the tungsten
sliding block is completely closed. If the gauge base opening is not
completely closed by the sliding block, the sliding block may
require cleaning. Refer to Appendix C for cleaning instructions.
Place the gauge over the test site. Release the gauge handle and
push it down until it is in the correct position. Ensure that the pin
engages the notch in the index rod.
Press
〈
START/ENTER
〉
.
After the count time has elapsed, the gauge displays the
measurement results in a series of six screens, as follows. Use the
up and down arrows to scroll through the various screens.
CAUTION
Do not store or transport the gauge unless the
tungsten sliding block is completely closed.
Increased radiation levels may violate transportation
regulations and cause excessive personnel exposure.
Depth: xx in.
Time: xxx sec.
G
AUG
E
O
P
E
RAT
IO
N
Содержание 3430-M
Страница 9: ......
Страница 14: ...xii NOTES...
Страница 19: ...Model 3430 1 5 Figure 1 1 Model 3430 Gauge and Accessories INTRODUCTION...
Страница 22: ...1 8 NOTES...
Страница 25: ...Model 3430 2 3 Figure 2 1 Direct Transmission Geometry Figure 2 2 Backscatter Geometry OPERATION THEORY...
Страница 26: ...2 4 Figure 2 3 Backscatter Surface Density Effects Top Layer Effect Curves...
Страница 28: ...2 6 Figure 2 4 Effect of Moisture on Depth of Measurement...
Страница 48: ...3 20 NOTES...
Страница 62: ...4 14 NOTES...
Страница 66: ...5 20 D xxxx M xxxx Use keys Moist R xxxx to view data Moist Avg xxxx to view data...
Страница 88: ...A 14 NOTES...
Страница 111: ...Model 3430 C 17 Figure C 2 3430 Gauge Assembly MAINTENANCE...
Страница 113: ...Model 3430 C 19 Figure C 3 3430 Gauge Base Mechanical Assembly Figure C 4 3430 Source Rod Handle Assembly MAINTENANCE...
Страница 115: ...Model 3430 C 21 Figure C 5 3430 Gauge Preamplifier Assembly MAINTENANCE...
Страница 117: ...Model 3430 C 23 Figure C 6 3430 Gauge Scaler Assembly MAINTENANCE...
Страница 128: ...E 2 Date MS DS Date MS DS STANDARD COUNT LOG Gauge Serial Number...
Страница 129: ...Model 3430 E 3 Date MS DS Date MS DS STANDARD COUNT LOG Gauge Serial Number STANDARD COUNT LOG...
Страница 130: ...E 4 Date MS DS Date MS DS STANDARD COUNT LOG Gauge Serial Number...
Страница 140: ...Index 8 NOTES...
Страница 141: ...Model 3430 Warranty 1 NOTES WARRANTY...