Model 2701-B
2–15
Target
The Model 2701-B can store up to four
Gmb
(Marshall) and
four
Gmm
(Voidless density) target values. A Gmb value is the
bulk density of a material, often determined using the
Marshall method. A Gmm value is the maximum theoretical
density for a test material as obtained in laboratory tests.
The gauge uses target values to determine the percent
compaction and/or percent voids (if a Gmm value is entered)
as compared to the measured density of the test material.
With the
Target
function, the operator can enter a new target
value, enable a stored target value, or disable a target value.
To access the
Target
function, press the
〈
TARGET
〉
key. The
gauge displays the
Target
menu:
To enter, enable, or disable a value, press the corresponding
numeric key. For example, to edit the Gmb (Marshall) value,
press
〈
1
〉
. The gauge displays the menu shown below. The
Gmm(Void) Value:
menu is similar and operates the same.
The target value menu shows the four stored target values (if
any) and the
New
and
Disable
options.
Gmb( MA) Val ue:
1. #. #
2. #. #
3. #. #
4. #. #
5. New
6. Di sabl e
- Tar get -
1. Gmb( Mar shal l )
2. Gmm( Voi dl ess)
Pr ess # t o Sel ect
2. G
ETTI
NG
S
TA
RTE
D