Fault Codes
The following codes relate to on-screen warnings that the Air XD will display when a fault is
encountered during normal operations.
Code
Fault Name
Fault Description
Fault Check
0x0009
Failed to Load
Settings
Failed to load and apply user
configured settings. As standard
the Air XD default settings will be
applied.
Re-apply custom settings
and check if unit saves and
restores files.
Contact Trolex or approved
distributor if this fails.
0x0301
No Sensor Comms
Communications with the
particulate sensor assembly has
been lost.
Check ‘D’ type connector to
sensor housing and power
cycle unit.
0x0304
Fan Current Too High
Particulate sampling fan may
have malfunctioned.
Check instrument is
operating within rated
temperature range.
0x0305
No Fan Current
Particulate sampling fan may
have malfunctioned.
Check instrument is
operating within rated
temperature range.
0x0306
Sensor Laser Current
High
Particulate sensor may have
failed or operating at excessive
temperatures.
Check instrument is
operating within rated
temperature range.
0x0307
Sensor Laser Current
Low
Particulate sensor may have
failed.
Contact Trolex or approved
distributor.
0x0308
Sensor Temp Too
High
Instrument is operating above
specified temperature limits. Unit
has gone into safe mode.
Reduce operating
temperature below 50°C
0x0402
Lost Memory Module
Comms
Communications to memory
module has failed. Data logging
functionality is lost.
Power cycle to reset
instrument processor and
retry.
Contact Trolex or approved
distributor if this fails.